McKinley C A, Parker P A
Institute of Biomedical Engineering, University of New Brunswick, Fredericton, Canada.
IEEE Trans Biomed Eng. 1991 Apr;38(4):379-82. doi: 10.1109/10.133234.
Evoked potentials (EP) contain information about various physiological parameters and the estimation and detection of these signals can aid in the diagnosis of many pathological conditions. However, the signal-to-noise ratio (SNR) for EP measurement is often very low, and thus signal processing techniques must be employed to enhance the SNR. A delay and sum beamformer acquisition system has the potential for significant SNR improvement in EP measurements. In this communication it is shown that an electrode array acquisition system implements a uniform coherent delay and sum beamformer. The performance of the beamformer is characterized in terms of the number of electrodes, and cross-channel correlation. When compared to conventional ensemble averaging, the beamformer reduces the number of response repetitions required to achieve a given SNR by a factor which approaches the number of channels in the acquisition system.