Matlaga Brian R, Kawamoto Satomi, Fishman Elliot
The Russell H Morgan Department of Radiology and Radiological Science, James Buchanan Brady Urological Institute, The Johns Hopkins University School of Medicine, Baltimore, Maryland 21209, USA.
Urology. 2008 Nov;72(5):1164-8. doi: 10.1016/j.urology.2008.03.051. Epub 2008 Jul 10.
Dual source computed tomography (CT), a novel technology that employs two different x-ray sources, may provide an image resolution that has not been possible with conventional single source CT. We tested the ability of dual source CT imaging to distinguish calcium oxalate (CaOx) stones from calcium phosphate (CaP) stones, and both types of calciferous stones from uric acid (UA) stones.
CaOx, CaP, and UA stones were placed in a tissue phantom and imaged with dual source multidetector CT at both 80kV and 140 kV. Hounsfield units (HU) of each stone were recorded for the 140 kV and the 80 kV datasets by drawing regions of interest by circle, hand-drawing, and using a volume calculation method. The differences between, and the ratios of, the HU obtained by the two energy sources were compared among the stone groups.
Between CaOx and CaP stones, HU differences (80kV-140kV difference, hand-drawing technique: CaOx = 270.7; CaP = 369.5; UA = 11.45; P = 0.020 for CaOx vs. CaP; P < 0.001 for CaOx and CaP vs. UA) as well as ratios (140kV:80kV ratio, hand-drawing technique: CaOx = 1.44; CaP = 1.51; UA = 1.04; P = 0.001 for CaOx vs. CaP; P < 0.001 for CaOx and CaP vs. UA) were significantly different. There were no significant differences in stone size among the groups.
Dual source CT has the ability to distinguish not only between calciferous stones and uric acid stones, but also among stones composed of different calcium salts. Further studies are warranted to confirm these findings in a clinical setting.
双源计算机断层扫描(CT)是一种采用两个不同X射线源的新技术,它可能提供传统单源CT无法实现的图像分辨率。我们测试了双源CT成像区分草酸钙(CaOx)结石与磷酸钙(CaP)结石,以及这两种含钙结石与尿酸(UA)结石的能力。
将CaOx、CaP和UA结石置于组织模型中,使用双源多探测器CT在80kV和140kV下进行成像。通过圆形绘制感兴趣区域、手动绘制以及使用体积计算方法,记录140kV和80kV数据集下每颗结石的亨氏单位(HU)。比较各结石组两种能量源获得的HU之间的差异及其比值。
在CaOx和CaP结石之间,HU差异(80kV - 140kV差值,手动绘制技术:CaOx = 270.7;CaP = 369.5;UA = 11.45;CaOx与CaP相比,P = 0.020;CaOx和CaP与UA相比,P < 0.001)以及比值(140kV:80kV比值,手动绘制技术:CaOx = 1.44;CaP = 1.51;UA = 1.04;CaOx与CaP相比,P = 0.001;CaOx和CaP与UA相比,P < 0.001)有显著差异。各组结石大小无显著差异。
双源CT不仅能够区分含钙结石与尿酸结石,还能区分由不同钙盐组成的结石。有必要进行进一步研究以在临床环境中证实这些发现。