一种用于辅助腰椎椎间融合的碳纤维植入物。力学测试。

A carbon fiber implant to aid interbody lumbar fusion. Mechanical testing.

作者信息

Brantigan J W, Steffee A D, Geiger J M

机构信息

Cleveland Spine and Arthritis Center, Ohio.

出版信息

Spine (Phila Pa 1976). 1991 Jun;16(6 Suppl):S277-82. doi: 10.1097/00007632-199106001-00020.

Abstract

A carbon-fiber-reinforced polymer implant has been designed to aid interbody lumbar fusion. The cage-like implant has ridges or teeth to resist pullout or retropulsion, struts to support weight bearing, and a hollow center for packing of autologous bone graft. Because carbon is radiolucent, bony healing can be imaged by standard radiographic techniques. The device has been mechanically tested in cadaver spines and compared with posterior lumbar interbody fusion performed with donor bone. The carbon device required a pullout force of 353 N compared with 126 N for donor bone. In compression testing, posterior lumbar interbody fusion performed with the carbon device bore a load of 5,288 N before failure of the vertebral bone. Posterior lumbar interbody fusion performed with donor bone failed at 4,628 N, and unmodified motion segments failed at 6,043 N. The carbon fiber implant separates the mechanical and biologic functions of posterior lumbar interbody fusion.

摘要

一种碳纤维增强聚合物植入物已被设计用于辅助腰椎椎间融合。这种笼状植入物有脊或齿以抵抗拔出或后移,有支柱以支撑负重,还有一个中空中心用于填充自体骨移植材料。由于碳是射线可透过的,骨愈合情况可通过标准放射技术成像。该装置已在尸体脊柱上进行了力学测试,并与使用异体骨进行的后路腰椎椎间融合术进行了比较。与异体骨所需的126 N拔出力相比,碳纤维装置需要353 N的拔出力。在压缩测试中,使用碳纤维装置进行的后路腰椎椎间融合术在椎体骨失效前承受了5288 N的负荷。使用异体骨进行的后路腰椎椎间融合术在4628 N时失效,未改良的运动节段在6043 N时失效。碳纤维植入物将后路腰椎椎间融合术的机械和生物学功能分开。