Habib Maged S, Lowell James A, Holliman Nick S, Hunter Andrew, Vaideanu Daniella, Hildreth Anthony, Steel David Hw
Ophthalmology Department, Sunderland Eye Infirmary, Sunderland, UK.
BMC Ophthalmol. 2008 Jul 23;8:13. doi: 10.1186/1471-2415-8-13.
Stereoscopic assessment of the optic disc morphology is an important part of the care of patients with glaucoma. The aim of this study was to assess stereoviewing of stereoscopic optic disc images using an example of the new technology of autostereoscopic screens compared to the liquid shutter goggles.
Independent assessment of glaucomatous disc characteristics and measurement of optic disc and cup parameters whilst using either an autostereoscopic screen or liquid crystal shutter goggles synchronized with a view switching display. The main outcome measures were inter-modality agreements between the two used modalities as evaluated by the weighted kappa test and Bland Altman plots.
Inter-modality agreement for measuring optic disc parameters was good [Average kappa coefficient for vertical Cup/Disc ratio was 0.78 (95% CI 0.62-0.91) and 0.81 (95% CI 0.6-0.92) for observer 1 and 2 respectively]. Agreement between modalities for assessing optic disc characteristics for glaucoma on a five-point scale was very good with a kappa value of 0.97.
This study compared two different methods of stereo viewing. The results of assessment of the different optic disc and cup parameters were comparable using an example of the newly developing autostereoscopic display technologies as compared to the shutter goggles system used. The Inter-modality agreement was high. This new technology carries potential clinical usability benefits in different areas of ophthalmic practice.
对视盘形态进行立体评估是青光眼患者护理的重要组成部分。本研究旨在以自动立体屏幕这一新技术为例,与液晶快门眼镜相比,评估立体视盘图像的立体观察效果。
在使用自动立体屏幕或与视图切换显示器同步的液晶快门眼镜时,独立评估青光眼视盘特征并测量视盘和视杯参数。主要观察指标是通过加权kappa检验和布兰德-奥特曼图评估的两种使用方式之间的模态间一致性。
测量视盘参数的模态间一致性良好[观察者1和观察者2的垂直杯盘比平均kappa系数分别为0.78(95%可信区间0.62 - 0.91)和0.81(95%可信区间0.6 - 0.92)]。在五分制量表上评估青光眼视盘特征的两种方式之间的一致性非常好,kappa值为0.97。
本研究比较了两种不同的立体观察方法。与所使用的快门眼镜系统相比,以新开发的自动立体显示技术为例,对视盘和视杯不同参数的评估结果具有可比性。模态间一致性较高。这项新技术在眼科实践的不同领域具有潜在的临床应用优势。