Optic disc imaging with spectral-domain optical coherence tomography: variability and agreement study with Heidelberg retinal tomograph.

机构信息

Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, PR China.

出版信息

Ophthalmology. 2012 Sep;119(9):1852-7. doi: 10.1016/j.ophtha.2012.02.033. Epub 2012 May 8.

Abstract

OBJECTIVE

To evaluate the agreement of optic disc measurements obtained with the Cirrus high-density optical coherence tomography (HD-OCT) and the Heidelberg retina tomograph (HRT) and compare the intervisit, test-retest variability between the instruments.

DESIGN

Prospective, cross-sectional study.

PARTICIPANTS

Two hundred seven subjects (109 glaucoma and 98 normal subjects).

METHODS

One eye from each individual was selected randomly for optic disc imaging by the Cirrus HD-OCT and the HRT. Areas of the optic disc and the cup, cup volume, vertical cup-to-disc ratio and cup-to-disc area ratio were compared between the instruments. The OCT measurements were corrected for ocular magnification using the Littman's formula. The measurement agreement was evaluated with the Bland-Altman plots. The intervisit test-retest variability was examined in 17 randomly selected glaucoma patients who underwent optic disc imaging weekly for 8 consecutive weeks. The intraclass correlation coefficients (ICC) and the reproducibility coefficients of the optic disc parameters were computed.

MAIN OUTCOME MEASURES

Measurement agreement, reproducibility coefficients, and ICCs of optic disc parameters.

RESULTS

The OCT measured smaller optic disc and rim areas and greater cup volume, vertical cup-to-disc ratio and cup-to-disc area ratio than the HRT did (all with P<0.001). There were proportional biases in the Bland-Altman plots between OCT and HRT optic disc measurements except for rim area and cup-to-disc area ratio. The 95% limits of agreement of rim area ranged between -0.28 and 0.88 mm(2) before, and between -0.22 and 0.92 mm(2) after correction for ocular magnification. Both OCT and HRT showed high test-retest reproducibility with ICCs ≥ 0.921. Although the reproducibility coefficient of OCT rim area (0.093 mm(2); 95% confidence interval [CI], 0.081-0.105 mm(2)) was significantly smaller than that of the HRT (0.186 mm(2); 95% CI, 0.163-0.210 mm(2); P = .018), there were no differences in the ICCs between the instruments.

CONCLUSIONS

Optic disc assessment by spectral-domain OCT and confocal scanning laser ophthalmoscopy demonstrates poor agreement but similarly low test-retest variability. The source of their disagreement and its effects on the detection of progression require further study.

摘要

目的

评估 Cirrus 高分辨率光学相干断层扫描(HD-OCT)和 Heidelberg 视网膜断层扫描仪(HRT)获得的视盘测量值的一致性,并比较两种仪器的随访间和复测间变异性。

设计

前瞻性、横断面研究。

参与者

207 名受试者(109 名青光眼患者和 98 名正常受试者)。

方法

从每位个体的一只眼随机选择进行 Cirrus HD-OCT 和 HRT 的视盘成像。比较两种仪器对视盘和杯的面积、杯容积、垂直杯盘比和杯盘面积比的测量值。使用 Littman 公式校正眼放大率对 OCT 测量值进行校正。使用 Bland-Altman 图评估测量值的一致性。对 17 名随机选择的青光眼患者进行随访,每周进行 8 次连续的视盘成像,以检查复测间的测试-复测变异性。计算视盘参数的组内相关系数(ICC)和可重复性系数。

主要观察指标

测量值一致性、视盘参数的可重复性系数和 ICC。

结果

OCT 测量的视盘和边缘面积较小,杯容积、垂直杯盘比和杯盘面积比较大,均高于 HRT(均 P<0.001)。除了边缘面积和杯盘面积比之外,OCT 和 HRT 的视盘测量值在 Bland-Altman 图中存在比例偏差。在未校正眼放大率之前,边缘面积的 95%一致性界限在-0.28 至 0.88mm²之间,校正后为-0.22 至 0.92mm²。OCT 和 HRT 的 ICC 均≥0.921,具有很高的复测间可重复性。尽管 OCT 边缘面积的可重复性系数(0.093mm²;95%置信区间 [CI],0.081-0.105mm²)明显小于 HRT(0.186mm²;95%CI,0.163-0.210mm²;P=0.018),但两种仪器的 ICC 无差异。

结论

频域光学相干断层扫描和共焦激光扫描检眼镜对视盘的评估显示出较差的一致性,但具有相似的低复测间变异性。需要进一步研究它们不一致的原因及其对视盘进展检测的影响。