Brumbaugh W G, Walther M J
U.S. Fish and Wildlife Service, National Fisheries Contaminant Research Center, Columbia, MO 65201.
J Assoc Off Anal Chem. 1989 May-Jun;72(3):484-6.
A combined wet chemical and dry ash digestion and use of a continuous-flow hydride generator coupled with a flame-heated quartz cell enabled the simple, precise, and highly automated atomic absorption determination of arsenic and selenium in tissues of whole fish. Percent relative standard deviation averaged 4% for each element; method detection limits (micrograms/g dry wt) were about 0.06 for arsenic and 0.04 for selenium. Digestion of samples proceeded with little operator attention and without perchloric acid. Analysis for arsenic as As(V) simplified sample preparation but care had to be exercised to avoid interferences from high concentrations of selenium.