Suppr超能文献

采用连续流动氢化物发生原子吸收分光光度法测定全鱼中的砷和硒。

Determination of arsenic and selenium in whole fish by continuous-flow hydride generation atomic absorption spectrophotometry.

作者信息

Brumbaugh W G, Walther M J

机构信息

U.S. Fish and Wildlife Service, National Fisheries Contaminant Research Center, Columbia, MO 65201.

出版信息

J Assoc Off Anal Chem. 1989 May-Jun;72(3):484-6.

PMID:2745375
Abstract

A combined wet chemical and dry ash digestion and use of a continuous-flow hydride generator coupled with a flame-heated quartz cell enabled the simple, precise, and highly automated atomic absorption determination of arsenic and selenium in tissues of whole fish. Percent relative standard deviation averaged 4% for each element; method detection limits (micrograms/g dry wt) were about 0.06 for arsenic and 0.04 for selenium. Digestion of samples proceeded with little operator attention and without perchloric acid. Analysis for arsenic as As(V) simplified sample preparation but care had to be exercised to avoid interferences from high concentrations of selenium.

摘要

采用湿化学法与干灰化法相结合,并使用与火焰加热石英池联用的连续流动氢化物发生器,可对全鱼组织中的砷和硒进行简单、精确且高度自动化的原子吸收测定。每种元素的相对标准偏差平均为4%;方法检测限(微克/克干重)砷约为0.06,硒约为0.04。样品消解过程几乎无需操作人员关注,且无需使用高氯酸。以五价砷形式分析砷简化了样品制备,但必须注意避免高浓度硒的干扰。