The characteristics of Fuji IP Cassette Type PII and application for radiation oncology quality assurance tests and portal imaging.

作者信息

Soh H S, Ung N M, Ng K H

机构信息

Department of Biomedical Imaging, University of Malaya, Kuala Lumpur, Malaysia.

出版信息

Australas Phys Eng Sci Med. 2008 Jun;31(2):146-50. doi: 10.1007/BF03178589.

Abstract

The advancement of digital imaging has prompted more medical institutions to go filmless. The computed radiography (CR) system is becoming an important tool not only in diagnostic imaging, but also in radiation oncology. A new CR system that was specially designed for the use in radiation oncology, Fuji IP cassette type PII has been introduced to the market in the middle of year 2006. This project aimed to study some basic physical characteristics of this new type of cassette and explore its application for performing quality assurance (QA) tests and portal imaging in radiotherapy. All the images were read by FCR 5000 Plus reader. The image was found to reach its saturation value of 1023 (due to the image was stored in 10 bits data) by depending on the sensitivity value being adjusted. The uniformity test gave the result of 0.12%. The cassette was used to perform the QA tests which were previously performed using film. All the results met the specification as stated in AAPM Task Group 40. The comparison for the portal images of PortalVision contrast-detail phantom showed that the spatial resolution of the images obtained by CR system (Fujifilm Co., Ltd., Tokyo, Japan) were better than the EPID (Varian Medical Systems, Inc., Palo Alto, USA) and film system (Eastman Kodak Co., New York, USA). The IP cassette type PII was found to be suitable as an alternative QA test tool and portal imaging in radiotherapy.

摘要

数字成像技术的进步促使更多医疗机构实现无胶片化。计算机X线摄影(CR)系统不仅在诊断成像中,而且在放射肿瘤学中都正成为一种重要工具。一种专门为放射肿瘤学使用而设计的新型CR系统——富士IP II型暗盒已在2006年年中推向市场。本项目旨在研究这种新型暗盒的一些基本物理特性,并探索其在放射治疗中进行质量保证(QA)测试和射野成像的应用。所有图像均由FCR 5000 Plus阅读器读取。通过调整灵敏度值,发现图像达到其1023的饱和值(由于图像存储为10位数据)。均匀性测试结果为0.12%。该暗盒用于执行以前使用胶片进行的QA测试。所有结果均符合美国医学物理学家协会任务组40中规定的规格。对PortalVision对比细节模体的射野图像进行比较表明,CR系统(日本东京富士胶片公司)获得的图像空间分辨率优于电子射野影像装置(美国帕洛阿尔托瓦里安医疗系统公司)和胶片系统(美国纽约伊士曼柯达公司)。发现IP II型暗盒适合作为放射治疗中QA测试和射野成像的替代工具。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验