Pesznyák Csilla, Fekete Gábor, Mózes Arpád, Kiss Balázs, Király Réka, Polgár István, Zaránd Pál, Mayer Arpád
Institute of Oncoradiology, Uzsoki Hospital, Budapest, Hungary.
Strahlenther Onkol. 2009 Jan;185(1):56-60. doi: 10.1007/s00066-009-1905-z. Epub 2009 Feb 18.
Quality assurance (QA) and quality control (QC) of different electronic portal imaging devices (EPID) and portal images with the PTW EPID QC PHANTOM.
Characteristic properties of images of different file formats were measured on Siemens OptiVue500aSi, Siemens BeamView Plus, Elekta iView, and Varian PortalVision and analyzed with the epidSoft 2.0 program in four radiation therapy centers. The portal images were taken with Kodak X-OMAT V and the Kodak Portal Localisation ReadyPack films and evaluated with the same program.
The optimal exposition both for EPIDs and portal films of different kind was determined. For double exposition, the 2+1 MU values can be recommended in the case of Siemens OptiVue500aSi Elekta iView and Kodak Portal Localisation ReadyPack films, while for Siemens BeamView Plus, Varian PortalVision and Kodak X-OMAT V film 7+7 MU is recommended.
The PTW EPID QC PHANTOM can be used not only for amorphous silicon EPIDs but also for images taken with a video-based system or by using an ionization chamber matrix or for portal film. For analysis of QC tests, a standardized format (used at the acceptance test) should be applied, as the results are dependent on the file format used.
使用PTW电子射野影像装置质量控制模体对不同的电子射野影像装置(EPID)和射野图像进行质量保证(QA)和质量控制(QC)。
在四个放射治疗中心,对西门子OptiVue500aSi、西门子BeamView Plus、医科达iView和瓦里安PortalVision上不同文件格式图像的特性进行测量,并使用epidSoft 2.0程序进行分析。射野图像采用柯达X-OMAT V和柯达射野定位成套胶片拍摄,并使用同一程序进行评估。
确定了不同类型的EPID和射野胶片的最佳曝光量。对于双重曝光,在西门子OptiVue500aSi、医科达iView和柯达射野定位成套胶片的情况下,推荐2 + 1MU值,而对于西门子BeamView Plus、瓦里安PortalVision和柯达X-OMAT V胶片,推荐7 + 7MU。
PTW EPID QC模体不仅可用于非晶硅EPID,还可用于基于视频系统拍摄的图像、使用电离室矩阵拍摄的图像或射野胶片。由于结果取决于所使用的文件格式,因此在质量控制测试分析中应采用标准化格式(验收测试时使用)。