Abstract

Electroporation is a long-established technique used to deliver molecules to cells. Most in vivo electroporation protocols entail applying square-wave, or monotonically-decreasing pulses but relatively few have explored the use of pulsed ac fields. This study measures the efficiency of electroporation in human kidney embryonal cells, using pulsed ac electric fields of peak amplitude 30-200 kV m(-1). The results indicate that optimum electroporation efficiencies of up to 70% can be achieved using pulses at frequencies of 20-160 kHz. Increasing the field strength results in higher electroporation efficiency, but also increases cell kill. This study confirms that efficient electroporation may be achieved using pulsed ac fields. This finding raises the possibility of a wider range of clinical and laboratory applications based on ac technology and avoiding the use of invasive needle electrodes.

摘要

电穿孔是一种长期使用的将分子导入细胞的技术。大多数体内电穿孔方案需要施加方波或单调递减脉冲,但相对较少有人探索使用脉冲交流电场。本研究使用峰值幅度为30 - 200 kV m(-1)的脉冲交流电场测量人肾胚胎细胞中的电穿孔效率。结果表明,使用20 - 160 kHz频率的脉冲可实现高达70%的最佳电穿孔效率。增加场强会导致更高的电穿孔效率,但也会增加细胞杀伤率。本研究证实使用脉冲交流电场可以实现高效电穿孔。这一发现增加了基于交流技术进行更广泛的临床和实验室应用的可能性,并避免使用侵入性针电极。