Mol A, Balasundaram A
Department of Diagnostic Sciences and General Dentistry, School of Dentistry, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599-7450, USA.
Dentomaxillofac Radiol. 2008 Sep;37(6):319-24. doi: 10.1259/dmfr/26475758.
To assess the accuracy of NewTom 9000 cone beam CT (CBCT) images for the detection and quantification of periodontal bone defects in three dimensions.
A sample of 146 sites in 5 dry skulls provided the ground truth (GT). Half of the sample had bone loss of at least 3 mm. Two metal spheres at each site ensured correspondence between GT and CBCT measurements. Skulls were submerged in water and scanned with the NewTom QR-DVT-9000. A full mouth series (FMX) was obtained of each skull using photostimulable phosphor plates. Six observers measured the bone height of each site and rated the presence or absence of bone loss. Measurements were compared to GT and A(z)-values were calculated from receiver operating characteristic curves.
The A(z)-value for CBCT was 0.74 (standard deviation (SD) = 0.14) and for FMX 0.48 (SD = 0.09). The difference was significant (ANOVA: P < 0.01). The diagnostic accuracy of CBCT was lower for anterior teeth (A(z) = 0.59) than for molars (A(z) = 0.82) and premolars (A(z) = 0.79) (Tukey's HSD (honestly significant difference): P < 0.01). The mean absolute difference between CBCT and GT was 1.27 mm (SD = 1.43) and between FMX and GT 1.49 mm (SE = 1.24) (ANOVA: P < 0.01). Measurements in the anterior mandible were less accurate than in other areas (Tukey's HSD: P < 0.01).
The NewTom 9000 cone beam CT scanner provides better diagnostic and quantitative information on periodontal bone levels in three dimensions than conventional radiography. The accuracy in the anterior aspect of the jaws is limited.
评估NewTom 9000锥形束CT(CBCT)图像在三维检测和定量分析牙周骨缺损方面的准确性。
从5个干燥颅骨上选取146个部位作为标准对照(GT)。样本中一半部位存在至少3mm的骨质流失。每个部位放置两个金属球以确保GT与CBCT测量结果的对应性。将颅骨浸入水中,使用NewTom QR-DVT-9000进行扫描。使用光激励荧光板获取每个颅骨的全口系列(FMX)图像。6名观察者测量每个部位的骨高度,并对骨质流失的有无进行评级。将测量结果与GT进行比较,并根据受试者工作特征曲线计算A(z)值。
CBCT的A(z)值为0.74(标准差(SD)=0.14),FMX的A(z)值为0.48(SD = 0.09)。差异具有统计学意义(方差分析:P < 0.01)。CBCT在前牙的诊断准确性(A(z)=0.59)低于磨牙(A(z)=0.82)和前磨牙(A(z)=0.79)(Tukey's HSD(真实显著差异):P < 0.01)。CBCT与GT之间的平均绝对差异为1.27mm(SD = 1.43),FMX与GT之间为1.49mm(SE = 1.24)(方差分析:P < 0.01)。下颌前部的测量准确性低于其他区域(Tukey's HSD:P < 0.01)。
与传统放射成像相比,NewTom 9000锥形束CT扫描仪在三维牙周骨水平的诊断和定量信息方面表现更佳。颌骨前部的准确性有限。