Suppr超能文献

免疫测定中嗜异性抗体的干扰:肌红蛋白值错误升高。

Interference by heterophilic antibodies in immunoassays: wrong increase of myoglobin values.

作者信息

Bonetti Antonio, Monica Cesare, Bonaguri Chiara, Gnocchi Cecilia, Russo Annalisa, Battistelli Luisita, Musiari Luisa, Pastori Paolo, Novarini Almerico

机构信息

Dept of Clinical Sciences, University of Parma, Italy.

出版信息

Acta Biomed. 2008 Aug;79(2):140-3.

Abstract

Aim of this work is to illustrate how analytical interference in immunoassay may produce serious errors in clinical laboratory results. The sophisticated quality assurance schemes used in many laboratories do not identify erroneous results arising from aberrant samples. Recently attention has been focused on the incidence and implication of false-positive results arising from the presence of certain substances in a patient's serum that interfere with one or more steps in immunoassays. In this paper, we present the case of a 92 year-old woman whose plasma myoglobin concentrations falsely increased when measured using the Beckman Access assay. We demonstrated that heterophilic antibodies accounted for the falsely increased myoglobin values, and we suggest how to resolve such situations.

摘要

这项工作的目的是说明免疫分析中的分析干扰如何在临床实验室结果中产生严重误差。许多实验室使用的复杂质量保证方案无法识别异常样本产生的错误结果。最近,注意力集中在患者血清中某些物质的存在干扰免疫分析中的一个或多个步骤而导致的假阳性结果的发生率及其影响。在本文中,我们介绍了一名92岁女性的病例,其血浆肌红蛋白浓度在使用贝克曼Access检测法测量时出现假性升高。我们证明了嗜异性抗体导致了肌红蛋白值的假性升高,并提出了如何解决此类情况的建议。