Moss Edward B, Trope Graham E
Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada.
J Glaucoma. 2008 Sep;17(6):489-93. doi: 10.1097/IJG.0b013e3181622532.
To measure closing pressure in a series of new Model FP7 silicone Ahmed Glaucoma Valves (AGVs), using an in vitro gravity-driven system, in which closing pressure was measured directly from a dual manometer/pressure reservoir.
A straight length of tubing served as an open manometer, which connected via a 3-way stopcock to the inlet of the AGV submerged under 1.5 cm of balanced saline solution (BSS). Six AGV were assessed in 5 sequential trials and the mean inlet pressure at 90 minutes was recorded as the closing pressure. After testing, a single valve was primed a second time using greater force and retested. Control trials were performed with a submerged 26-G cannula opening directly into the BSS bath.
At 90 minutes, the 6 valves equilibrated at a mean inlet pressure of 7.1 mm Hg, with a range from 1.4 to 13.5 mm Hg. Closing pressures at or less than 5 mm Hg were measured in half (3/6) of the valves tested. The average flow of BSS across the valves in the final 30 minutes was less than 1.5x10(-2) microL/min. In the single valve reperfused under greater pressure, the closing pressure increased.
Model FP7 AGVs, in vitro, exhibit significant variability in closing pressure, with half closing at intraocular pressures considered potentially problematic in clinical situations. The results of the high-pressure perfusion experiment suggest that more research into the priming process is required so a precise description can be developed for surgeons.
使用体外重力驱动系统测量一系列新型FP7型硅胶艾哈迈德青光眼引流阀(AGV)的关闭压力,在该系统中,关闭压力直接从双压力计/压力储器测量。
一段直管用作开放式压力计,通过三通旋塞连接到浸没在1.5厘米平衡盐溶液(BSS)下的AGV的入口。在5次连续试验中评估了6个AGV,并将90分钟时的平均入口压力记录为关闭压力。测试后,使用更大的力对单个阀门进行第二次灌注并重新测试。使用直接通向BSS浴槽的浸没式26G套管进行对照试验。
在90分钟时,6个阀门在平均入口压力7.1毫米汞柱处达到平衡,范围为1.4至13.5毫米汞柱。在测试的一半(3/6)阀门中测量到关闭压力等于或小于5毫米汞柱。在最后30分钟内通过阀门的BSS平均流量小于1.5×10(-2)微升/分钟。在更高压力下重新灌注的单个阀门中,关闭压力增加。
体外实验中,FP7型AGV的关闭压力存在显著差异,在临床情况下,一半阀门在被认为可能有问题的眼内压下关闭。高压灌注实验结果表明,需要对灌注过程进行更多研究,以便为外科医生制定精确的描述。