The relationship between increased hip range of motion, wear, and locking mechanism failure in the Harris-Galante acetabular component.

作者信息

Imai Hiroshi, Mashima Naohiko, Takahashi Toshiaki, Yamamoto Haruyasu

机构信息

Department of Bone and Joint Surgery, Ehime University Graduate School of Medicine, Ehime, Japan.

出版信息

J Arthroplasty. 2009 Sep;24(6):892-7. doi: 10.1016/j.arth.2008.07.003. Epub 2008 Oct 9.

Abstract

We performed both clinical and radiographic evaluations of 178 patients (190 hips) who had undergone cementless total hip arthroplasties using Harris-Galante I/II porous cups after an average 12-year follow-up period (range, 8-18 years). We revised 15 Harris-Galante I/II porous cups (7.8%), and the locking mechanism was broken in 10 revised cups (67%). There was a significant association between locking mechanism failure and linear polyethylene wear. We observed a significant positive correlation between linear polyethylene wear and increased ranges of motion such as flexion, adduction, and external rotation at the last follow-up visit after the primary operation. Increased ranges of motion seen in Asians induced higher linear polyethylene wear and locking mechanism failure due to impingement of the neck and cup.

摘要

我们对178例患者(190髋)进行了临床和影像学评估,这些患者均接受了使用Harris-Galante I/II多孔杯的非骨水泥型全髋关节置换术,平均随访期为12年(范围8 - 18年)。我们翻修了15个Harris-Galante I/II多孔杯(7.8%),其中10个翻修杯(67%)的锁定机制出现断裂。锁定机制失效与线性聚乙烯磨损之间存在显著关联。在初次手术后的最后一次随访中,我们观察到线性聚乙烯磨损与诸如屈曲、内收和外旋等活动范围增加之间存在显著正相关。亚洲人出现的活动范围增加由于颈部与髋臼杯的撞击导致了更高的线性聚乙烯磨损和锁定机制失效。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验