Mai Thê Anh, Lemij Hans G
Department of Ophthalmology, University Medical Center Groningen, Hanzeplein 1, 9700 RB Groningen, the Netherlands.
Arch Ophthalmol. 2008 Oct;126(10):1359-64. doi: 10.1001/archopht.126.10.1359.
To investigate the longitudinal corneal birefringence (corneal polarization axis [CPA] and corneal polarization magnitude [CPM]) variability in scanning laser polarimetry with variable corneal compensation and its effect on retinal nerve fiber layer measurements. Method We analyzed scanning laser polarimetry images obtained every 6 months for 3.2 years in 16 healthy eyes, 38 eyes with ocular hypertension, and 53 eyes with glaucoma in 107 white participants. Differences in values between each intraeye CPA and CPM measurement and the first measurement were used to investigate the variability and any trend with time, and any association with age or diagnosis. We also calculated the percentage of these values within the range of +/-5 degrees or +/-5 nm, respectively. Any effect of corneal birefringence variability on the retinal nerve fiber layer measurements was also evaluated.
The CPA and CPM measurement variability showed no trend with time and did not differ between diagnostic groups. It did not appear to be affected by age. With more than 90% of the CPA and CPM measurement variability within the range of +/-5 degrees or +/-5 nm, no significant effect on the retinal nerve fiber layer measurements was observed.
The CPA and CPM measurement variability did not differ between groups, showed no trend over time, was independent of subject age, and did not seem to systematically affect retinal nerve fiber layer reproducibility.
研究在可变角膜补偿的扫描激光偏振仪中纵向角膜双折射(角膜偏振轴[CPA]和角膜偏振幅度[CPM])的变异性及其对视网膜神经纤维层测量的影响。方法我们分析了107名白人参与者中16只健康眼睛、38只高眼压眼睛和53只青光眼眼睛在3.2年中每6个月获取的扫描激光偏振仪图像。每次眼内CPA和CPM测量值与首次测量值之间的差异用于研究变异性以及随时间的任何趋势,以及与年龄或诊断的任何关联。我们还分别计算了这些值在+/-5度或+/-5纳米范围内的百分比。还评估了角膜双折射变异性对视网膜神经纤维层测量的任何影响。
CPA和CPM测量变异性未显示随时间的趋势,诊断组之间也无差异。它似乎不受年龄影响。超过90%的CPA和CPM测量变异性在+/-5度或+/-5纳米范围内,未观察到对视网膜神经纤维层测量有显著影响。
CPA和CPM测量变异性在各组之间无差异,未显示随时间的趋势,与受试者年龄无关,并且似乎不会系统性地影响视网膜神经纤维层的可重复性。