Kim Sinyoung, Kim Jeong-Ho, Yoon Seoyoung, Park Youn-Hee, Kim Hyon-Suk
Department of Laboratory Medicine, Yonsei University College of Medicine, Seodaemun-gu, Seoul, Korea.
J Clin Microbiol. 2008 Dec;46(12):3919-23. doi: 10.1128/JCM.01603-08. Epub 2008 Oct 22.
Various automated chemiluminescence immunoassay (CLIA) analyzers for the detection of antibodies to hepatitis C virus (HCV) are now commercially available in clinical laboratories and are replacing conventional enzyme immunoassays. We investigated the performance of four anti-HCV CLIAs (the Architect Anti-HCV assay on the Architect i2000 system, the Vitros Anti-HCV assay on the Vitros ECiQ Immunodiagnostic System, the Access HCV Ab PLUS assay on the UniCel DxI 800 analyzer, and the newly developed Elecsys Anti-HCV assay on the Cobas e 411 analyzer). The total percent coefficient of variation values of imprecision were 3.5 to 5.7% with positive control materials and 7.2 to 10.2% with negative control materials. The agreement between the results of the Elecsys, Architect, Vitros, and Access CLIAs ranged from 94.5 to 98.1%. The clinical sensitivity of all CLIAs was 100%. Each CLIA showed excellent reproducibility and clinical sensitivity. The Elecsys, Architect, Vitros, and Access CLIAs showed clinical specificities of 98.2, 98.8, 96.5, and 98.2%.
目前临床实验室已有多种用于检测丙型肝炎病毒(HCV)抗体的自动化化学发光免疫分析(CLIA)分析仪,它们正在取代传统的酶免疫分析方法。我们研究了四种抗HCV CLIA的性能(在Architect i2000系统上的Architect抗HCV检测、在Vitros ECiQ免疫诊断系统上的Vitros抗HCV检测、在UniCel DxI 800分析仪上的Access HCV Ab PLUS检测以及在Cobas e 411分析仪上新开发的Elecsys抗HCV检测)。阳性对照材料的不精密度总变异系数百分比值为3.5%至5.7%,阴性对照材料为7.2%至10.2%。Elecsys、Architect、Vitros和Access CLIA检测结果之间的一致性范围为94.5%至98.1%。所有CLIA的临床敏感性均为100%。每种CLIA均显示出出色的重复性和临床敏感性。Elecsys、Architect、Vitros和Access CLIA的临床特异性分别为98.2%、98.8%、96.5%和98.2%。