Amplitude information from Doppler color flow mapping systems: a preliminary study of the power mode.

作者信息

Parro A, Philpot E F, Czuwala P J, Fan P H, Nanda N C, Cape E G, Yoganathan A P, Helmcke F, Fujii H

机构信息

Division of Cardiovascular Disease, University of Alabama, Birmingham 35294.

出版信息

J Am Coll Cardiol. 1991 Oct;18(4):997-1003. doi: 10.1016/0735-1097(91)90759-3.

Abstract

The flow of a saline-glycerin solution with sand particles through a continuous in vitro flow system was imaged by using two commercially available Doppler color flow mapping systems in a power mode (Toshiba SSH-160A and Advanced Technology Laboratories [ATL] Ultramark 9). The images generated from seven solutions with particle concentrations ranging from 0.0001 x 10(12) to 6 x 10(12) particles/liter and a mean velocity of 30 cm/s measured with use of pulsed Doppler ultrasound were used to examine the dependence of the power mode on particle concentration. To examine the velocity dependence, 20 mean velocities ranging from 0.1 to 0.53 m/s (3 to 30 liters/min) and three particle concentrations (1, 3 and 6 x 10(12) particles/liter) in the solution were used. The recorded images were digitized and analyzed off-line. The SUM values, or the adjusted color intensity levels in delineated areas of interest in the displayed flow, were compared. In general, the power mode was sensitive in displaying slower velocity flows; in the selected particle concentration and velocity ranges, it was both velocity and concentration dependent. The specific dependence differed for the two color flow mapping systems.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验