Von Bibra H, Stempfle H, Poll A, Schlief R, Emslander H
I. Medizinische Klinik of the Technical University of Munich, Germany.
Echocardiography. 1991 Sep;8(5):533-40. doi: 10.1111/j.1540-8175.1991.tb01014.x.
Flow detection by color Doppler is impaired by low velocity of flow and increasing attenuation and depth of ultrasound penetration. The effects of increased echogenicity on flow detection (Toshiba SSH 65A) were thus studied in a flow model, which yielded similar strengths of Doppler signals as seen in the clinical routine, by adding microbubble solutions to the blood analog fluid (45% aqueous glycerin) and comparing signal strength (score 0-5) prior to and after contrast. The flow within the plexiglass tube with less than 3 degrees angle of coincidence for Doppler interrogation was laminar with a parabolic velocity profile at physiological velocities and pressures. In comparison with various contrast agents at a flow velocity of 18 cm/sec and 19 dB attenuation, flow was not detectable in control color Doppler, after 4 mL of the blood analog fluid, and after 4 mL of an agitated saline solution 9 mg/mL, but visible after 4 mL of Echovist, a polysaccharide solution with reproducible bubble size and concentration and after 4 mL of an agitated polygelatin solution. Increasing concentration of Echovist (50-400 mg/mL) improved flow detection. Echovist 200 mg/mL enhanced the score by 1.1 + -0.6 for velocities 5-20 cm/sec (P less than 0.01), by 1.6 + -0.8 for 40-100 cm/sec (P less than 0.001), and by 1.1 + -0.6 for velocities greater than 150 cm/sec (P less than 0.001).(ABSTRACT TRUNCATED AT 250 WORDS)
彩色多普勒血流检测会受到血流速度低、超声穿透衰减增加及深度的影响。因此,在一个血流模型中研究了回声增强对血流检测(东芝SSH 65A)的影响,该模型通过向血液模拟液(45%的甘油水溶液)中添加微泡溶液,并比较造影前后的信号强度(0 - 5分),得出与临床常规中所见相似强度的多普勒信号。在有机玻璃管内,多普勒询问的夹角小于3度时,在生理速度和压力下,血流呈层流,速度分布呈抛物线状。在流速为18 cm/秒、衰减为19 dB的情况下,与各种造影剂相比,在对照彩色多普勒中,注入4 mL血液模拟液和4 mL 9 mg/mL的搅动盐水溶液后,血流无法检测到,但注入4 mL Echovist(一种具有可重复气泡大小和浓度的多糖溶液)和4 mL搅动的聚明胶溶液后可见血流。增加Echovist的浓度(50 - 400 mg/mL)可改善血流检测。Echovist 200 mg/mL使5 - 20 cm/秒速度下的评分提高了1.1±0.6(P小于0.01),40 - 100 cm/秒速度下提高了1.6±0.8(P小于0.001)大于150 cm/秒速度下提高了1.1±0.6(P小于0.001)。(摘要截选至250字)