Instituto de Quimica, Universidade de São Paulo, Av. Prof. Lineu Prestes, 748-05508-900, São Paulo, SP, Brazil.
Talanta. 2001 Apr 12;54(2):403-9. doi: 10.1016/s0039-9140(01)00311-3.
A polarographic DC(tast) method with the static mercury drop electrode, SMDE, was developed for determination of the flotation collector ethyl xanthate (EtX) in the concentration range from 1x10(-5) to 8x10(-5) M. The potentiality of the method was demonstrated by evaluating the capacity of powdered galena ore (PbS) to adsorb EtX in a titration-like procedure. Sulfide could be determined simultaneously with EtX because in NaOH electrolyte their anodic waves are well separated (E(1/2) congruent with-0.72 and -0.42 V versus Ag/AgCl, respectively). In addition, a new FIA method was developed by adapting a simple device to the tip of the glass capillary of a mercury electrode and doing amperometric detection at a fixed potential of -0.1 V, always in the DC(tast) mode. No oxygen removal was required. Reproducible results were obtained at a frequency of 72 injections per h, with automatic renewal of the SMDE every second. The calibration curve for freshly prepared EtX standards rendered a straight line from 5x10(-6) to 8x10(-5) M with correlation coefficient of 0.997, suitable for real applications in flotation processes and its effluents.
采用静态汞滴电极(SMDE)的极谱直流(tast)法,用于测定浮选捕收剂乙基黄原酸盐(EtX)在 1x10(-5) 至 8x10(-5) M 的浓度范围内的浓度。通过评估粉状方铅矿(PbS)在类似于滴定的程序中吸附 EtX 的能力,证明了该方法的潜力。因为在 NaOH 电解质中,它们的阳极波很好地分离(相对于 Ag/AgCl 的 E(1/2)分别为-0.72 和-0.42 V),所以可以同时测定硫化物和 EtX。此外,通过在汞电极的玻璃毛细管尖端适配简单的装置并在固定的-0.1 V 电位下进行安培检测,在直流(tast)模式下,开发了一种新的 FIA 方法。不需要去除氧气。以每小时 72 次注射的频率获得重现性结果,每秒钟自动更新 SMDE。新制备的 EtX 标准品的校准曲线在 5x10(-6) 至 8x10(-5) M 的范围内呈直线,相关系数为 0.997,适用于浮选过程及其废水的实际应用。