Yi H, Wu K, Hu S, Cui D
Department of Chemistry, Wuhan University, Wuhan 430072, People's Republic of China; Department of Food Science, Hubei Agriculture College, Jingzhou 434103, People's Republic of China.
Talanta. 2001 Dec 24;55(6):1205-10. doi: 10.1016/s0039-9140(01)00531-8.
In this paper, a new voltammetric method for the determination of phenol is described. In pH 8.00 phosphate buffer and in the presence of long-chain cationic surfactant-cetyltrimethylammonium bromide-phenol has a very sensitive oxidation peak at 0.47 V (vs. SCE) on the Nafion-modified glassy carbon electrode (GCE). The experimental parameters, such as supporting electrolyte and pH values, amounts of Nafion, varieties and concentration of surfactants, accumulation potential and time, as well as scan rate were optimized. The peak current is linear with the concentration of phenol in the range from 8x10(-9) to 1x10(-5) M, and the detection limit is 1x10(-9) M after being accumulated at -0.50 V (vs. SCE) for 3 min. Trace levels of phenol in water samples were determined by using this voltammetric method, the average recovery was calculated to be 99.56%.
本文描述了一种测定苯酚的新伏安法。在pH 8.00的磷酸盐缓冲液中,在长链阳离子表面活性剂十六烷基三甲基溴化铵存在下,苯酚在Nafion修饰的玻碳电极(GCE)上于0.47 V(vs. SCE)处有一个非常灵敏的氧化峰。对诸如支持电解质和pH值、Nafion用量、表面活性剂的种类和浓度、富集电位和时间以及扫描速率等实验参数进行了优化。峰电流与苯酚浓度在8×10⁻⁹至1×10⁻⁵ M范围内呈线性关系,在 -0.50 V(vs. SCE)处富集3分钟后,检测限为1×10⁻⁹ M。采用该伏安法测定了水样中的痕量苯酚,计算得出平均回收率为99.56%。