Suppr超能文献

流动注射分析装置中直接化学发光法测定没食子酸

Determination of tannic acid by direct chemiluminescence in a FIA assembly.

机构信息

Departamento Quimica, Bioquimica y Biologia Molecular, University Cardenal Herrera-CEU, 46113 Moncada, Valencia, Spain.

出版信息

Talanta. 2002 Dec 6;58(6):1243-51. doi: 10.1016/s0039-9140(02)00410-1.

Abstract

The determination of tannic acid is performed in a FIA assembly on the basis of the analytical output obtained by oxidation of the acid. The analyte solution was daily prepared in a mixture of quinine as sensitiser and perchloric acid and it was injected into a pure water stream acting as a carrier. This solution merges with the mixture potassium permanganate in perchloric medium and the resulting chemiluminescence is monitored. The method was applied over the range 0.5-20 mg l(-1) of tannic acid with a LOD 100 mug l(-1). The reproducibility was 2.1% and the sample throughput 54 h(-1). The influence of foreign substances was studied and the new method is applied to the determination of tannic acid in pharmaceutical and galenic formulations in human urine and surface waters.

摘要

丹宁酸的测定是在 FIA 组件中进行的,基于酸氧化得到的分析结果。分析物溶液每天在奎宁作为敏化剂和高氯酸的混合物中制备,并注入作为载体的纯水流中。该溶液与高氯酸介质中的高锰酸钾混合物混合,监测产生的化学发光。该方法适用于丹宁酸 0.5-20mg/L 的浓度范围,检测限为 100 微克/L。重现性为 2.1%,样品通量为 54 小时/次。研究了杂质的影响,新方法应用于人尿和地表水中的药物和制剂中的丹宁酸的测定。