We investigated the ferrous sulfate-benzoic acid-xylenol orange (FBX) aqueous chemical dosimeter for measurement of virtual (dynamic) wedge profiles on a linear accelerator. The layout for irradiation of the FBX-filled tubes mimicked a conventional linear detector array geometry. A comparison of the resulting measurements with film-measured profiles showed that, in the main beam region, the difference between the FBX system and the film system was within +/-2% and that, in the penumbra region, the difference varied from +/-1 mm to +/-2.5 mm in terms of positional equivalence, depending on the size of the dosimeter tubes. We thus believe that the energy-independent FBX dosimetry system can measure virtual wedge profiles with reasonable accuracy at reasonable cost. However, efficiency improvement is required before this dosimetry system can be accepted into routine practice.

摘要

我们研究了用于在线性加速器上测量虚拟(动态)楔形野轮廓的硫酸亚铁-苯甲酸-二甲酚橙(FBX)水溶液化学剂量计。填充FBX的试管的照射布局模仿了传统的线性探测器阵列几何形状。将所得测量结果与胶片测量的轮廓进行比较表明,在主射野区域,FBX系统与胶片系统之间的差异在±2%以内;在半影区域,根据剂量计试管的大小,在位置等效方面,差异从±1毫米到±2.5毫米不等。因此,我们认为能量无关的FBX剂量测定系统能够以合理的成本进行合理准确的虚拟楔形野轮廓测量。然而,在该剂量测定系统能够被纳入常规实践之前,需要提高其效率。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/cf84/5722356/c3025f7848d5/ACM2-9-206-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验