Department of Radiation Oncology, Max Hospital, Saket, New Delhi 110017, India.
Phys Med. 2013 Jun;29(4):368-73. doi: 10.1016/j.ejmp.2012.05.005. Epub 2012 Jun 9.
The ferrous sulphate-benzoic acid-xylenol orange (FBX) chemical dosimeter, due to its aqueous form can measure average volume doses and hence may overcome the limitations of point dosimetry. The present study was undertaken to validate the use of FBX dosimeter for rectum and bladder dose measurement during intracavitary brachytherapy (ICBT) and transperineal interstitial brachytherapy (TIB). We filled cylindrical polypropylene tubes (PT) and Foley balloons (FB) with FBX solution and used them as substitutes for rectum and bladder dose measurements respectively. A water phantom was fabricated with provision to place the Fletcher-type ICBT and MUPIT template applicators, and FBX filled PT and FB within the phantom. The phantom was then CT scanned for treatment planning and subsequent irradiation. Our results show that the average difference between DVH derived dose value and FBX measured dose is 3.5% (PT) and 13.7% (FB) for ICBT, and 9% (PT) and 9.9% (FB) for TIB. We believe that the FBX system should be able to provide accuracy and precision sufficient for routine quality assurance purposes. The advantage of the FBX system is its water equivalent composition, average volume dose measuring capability, and energy and temperature independent response as compared to TLD or semiconductor dosimeters. However, detailed studies will be needed with regards to its safety before actual in-vivo dose measurements are possible with the FBX dosimeter.
硫酸亚铁-苯甲酸-二甲酚橙(FBX)化学剂量计,由于其水溶液形式可以测量平均体积剂量,因此可能克服点剂量测量的局限性。本研究旨在验证 FBX 剂量计在腔内近距离放射治疗(ICBT)和经会阴间隙近距离放射治疗(TIB)中直肠和膀胱剂量测量的适用性。我们将 FBX 溶液填充到圆柱形聚丙烯管(PT)和 Foley 球囊(FB)中,并分别将其用作直肠和膀胱剂量测量的替代品。制作了一个水模体,用于放置 Fletcher 型 ICBT 和 MUPIT 模板施源器,以及 FBX 填充的 PT 和 FB 置于模体中。然后对该模体进行 CT 扫描以进行治疗计划和后续照射。我们的结果表明,对于 ICBT,DVH 衍生剂量值与 FBX 测量剂量之间的平均差异为 3.5%(PT)和 13.7%(FB),对于 TIB,为 9%(PT)和 9.9%(FB)。我们相信,FBX 系统应该能够提供足够的准确性和精密度,以满足常规质量保证目的。与 TLD 或半导体剂量计相比,FBX 系统的优势在于其水当量组成、平均体积剂量测量能力以及能量和温度无关的响应。然而,在实际体内剂量测量之前,需要对 FBX 剂量计的安全性进行详细研究。