Elder P J, Coveney F M, Welsh A D
Medical Physics Department, Cheltenham General Hospital, Gloucestershire, UK.
Phys Med Biol. 1995 Apr;40(4):683-9. doi: 10.1088/0031-9155/40/4/013.
Three dosimetric methods for commissioning the dynamic wedges on a Varian 600C linear accelerator operating at 4.7 MV were compared. The techniques involved the use of ionization chambers, x-ray verification film and solid state detectors. Ionization chambers gave the most accurate results, as expected, but data acquisition was very time consuming. Film dosimetry was found to have good spatial resolution for beam profiles, but problems were encountered with the acquisition of depth doses. Solid state detectors generally showed reasonable agreement with the ionization chambers.
比较了三种用于在4.7MV运行的Varian 600C直线加速器上调试动态楔形板的剂量测定方法。这些技术涉及电离室、X射线验证胶片和固态探测器的使用。正如预期的那样,电离室给出了最准确的结果,但数据采集非常耗时。发现胶片剂量测定法对射野轮廓具有良好的空间分辨率,但在获取深度剂量时遇到了问题。固态探测器通常与电离室显示出合理的一致性。