Glasser Lewis, Murphy Carole A, Machan Jason T
Department of Pathology, Rhode Island Hospital, USA.
Am J Clin Pathol. 2009 Jan;131(1):58-63. doi: 10.1309/AJCPCHHTZ6UDQ6WF.
Although automation has improved the accuracy and precision of blood cell counts and is more rapid and less labor-intensive, cerebrospinal fluid (CSF) samples are still counted manually. We compared the IRIS iQ200 Body Fluids Module (Iris Diagnostics, Chatsworth, CA) and the Beckman-Coulter LH750 (Beckman-Coulter, Brea, CA) with manual counts and evaluated the impact of automation on the laboratory if clinically acceptable performance was to be maintained. Automated counts were compared with manual counts on 313 specimens. Clinical reliability was estimated using the weighted kappa coefficient and the impact of errors discussed in the context of a historic census of 3,653 samples spanning 19 months. Nucleated cell counts had a reliability of 0.73 for the LH750 and 0.84 for the iQ200. However, our results showed unacceptable rates of error at counts less than 200/microL (200 x 106/L) for the LH750 and less than 50/microL (50 x 106/L) for the iQ200, representing 94% and 83% of the census specimens, respectively. If clinical reliability is to be maintained, neither the LH750 nor iQ200 would have a significant impact on improving the efficiency of the laboratory because of the high percentage of low CSF cell counts.
尽管自动化提高了血细胞计数的准确性和精密度,且速度更快、劳动强度更低,但脑脊液(CSF)样本仍需手工计数。我们将IRIS iQ200体液模块(Iris Diagnostics,加利福尼亚州查茨沃思)和贝克曼库尔特LH750(贝克曼库尔特,加利福尼亚州布雷亚)与手工计数进行了比较,并评估了若要维持临床可接受的性能,自动化对实验室的影响。对313份标本的自动化计数与手工计数进行了比较。使用加权kappa系数估计临床可靠性,并在涵盖19个月的3653份样本的历史普查背景下讨论误差的影响。LH750的有核细胞计数可靠性为0.73,iQ200为0.84。然而,我们的结果显示,对于LH750,细胞计数低于200/微升(200×10⁶/升)时误差率不可接受,对于iQ200,低于50/微升(50×10⁶/升)时误差率不可接受,分别占普查标本的94%和83%。如果要维持临床可靠性,由于低脑脊液细胞计数的比例很高,LH750和iQ200对提高实验室效率都不会有显著影响。