Ziegler Ronny, Brendel Bernhard, Rinneberg Herbert, Nielsen Tim
Philips Research Europe-Hamburg, Röntgenstr. 24, 22335 Hamburg, Germany.
Phys Med Biol. 2009 Jan 21;54(2):413-31. doi: 10.1088/0031-9155/54/2/016. Epub 2008 Dec 19.
Using a statistical (chi-square) test on simulated data and a realistic noise model derived from the system's hardware we study the performance of diffuse optical tomography systems for fluorescence imaging. We compare the predicted smallest size of detectable lesions at various positions in slab and cup geometry and model how detection sensitivity depends on breast compression and lesion fluorescence contrast. Our investigation shows that lesion detection is limited by relative noise in slab geometry and by absolute noise in cup geometry.