Suppr超能文献

毛细管区带电泳轴向电容耦合非接触电导检测十年回顾

Ten years of axial capacitively coupled contactless conductivity detection for CZE--a review.

作者信息

Kubán Pavel, Hauser Peter C

机构信息

Institute of Analytical Chemistry of the Academy of Sciences of the Czech Republic, Veverí, Brno, Czech Republic.

出版信息

Electrophoresis. 2009 Jan;30(1):176-88. doi: 10.1002/elps.200800478.

Abstract

Contactless conductivity detection is universal for CE in that all charged species can be quantified, and it is particularly attractive for those inorganic and organic ions that are not directly accessible by optical means. It is not necessary to make precise alignments or to create an optical window into the capillary. Commercial detectors that can be retrofitted to existing instruments are available at a cost that is lower than that of any other electrophoresis detector. The approach is also suited for lab-on-chip devices. It is attempted in this review to summarize some of the expertise accumulated since the introduction of the axial contactless conductivity detector to CE 10 years ago.

摘要

非接触式电导检测对于毛细管电泳而言具有通用性,因为所有带电物质都可以被定量,对于那些无法通过光学手段直接检测的无机和有机离子尤其具有吸引力。无需进行精确校准或在毛细管中制造光学窗口。可以改装到现有仪器上的商用检测器价格低于任何其他电泳检测器。该方法也适用于芯片实验室设备。本文试图总结自10年前轴向非接触式电导检测器引入毛细管电泳以来积累的一些专业知识。