Wu Wen-Chau, Edlow Brian L, Elliot Mark A, Wang Jiongjiong, Detre John A
Department of Radiology, University of Pennsylvania,Philadelphia, PA 19104, USA.
IEEE Trans Med Imaging. 2009 May;28(5):703-9. doi: 10.1109/TMI.2008.2012020. Epub 2009 Jan 13.
The purpose of this study is to evaluate cardiac and respiratory modulations in the signals of arterial spin labeling (ASL) perfusion magnetic resonance imaging (MRI) using RETROICOR, an image domain based retrospective correction method. Systematic comparisons were conducted for tagging schemes, pulsed (PASL) versus frequency-modulated continuous (CASL) methods, and the use of background suppression (BGS). Results showed that cardiac pulsation accounted for more signal fluctuation in PASL than in CASL (two-tailed paired Student's t-test, p < 10(-3)), whereas no significant difference was found in the effect of respiratory motion ( p=0.55) on the two tagging schemes studied. For PASL, significantly more improvement was achieved by the inclusion of cardiac pulsation than respiratory motion in RETROICOR ( p < 0.01). On the contrary, the inclusion of respiratory motion offers more improvement for CASL ( p < 0.02). BGS effectively improved the temporal signal-to-noise ratio (tSNR) as previous studies reported, but no significant difference was measured in the spectral power of physiological modulations relative to the entire spectrum of PASL signals before and after the superimposition of BGS ( p=0.63 for cardiac component, p=0.67 for respiratory component). Thus, we conclude that BGS reduces noise without spectral selectivity, and the improvements of tSNR from RETROICOR and BGS are additive. CASL with a labeling duration at a multiple of an R-R interval can be used to minimize signal fluctuation originating from cardiac pulsation.
本研究的目的是使用RETROICOR(一种基于图像域的回顾性校正方法)评估动脉自旋标记(ASL)灌注磁共振成像(MRI)信号中的心脏和呼吸调制。对标记方案、脉冲式(PASL)与调频连续式(CASL)方法以及背景抑制(BGS)的使用进行了系统比较。结果表明,与CASL相比,心脏搏动在PASL中导致的信号波动更多(双侧配对学生t检验,p < 10(-3)),而在所研究的两种标记方案中,呼吸运动的影响没有显著差异(p = 0.55)。对于PASL,在RETROICOR中纳入心脏搏动比纳入呼吸运动带来的改善更显著(p < 0.01)。相反,对于CASL,纳入呼吸运动带来的改善更多(p < 0.02)。正如先前研究所报道的,BGS有效地提高了时间信噪比(tSNR),但在BGS叠加前后,相对于PASL信号的整个频谱,生理调制的频谱功率没有显著差异(心脏成分p = 0.63,呼吸成分p = 0.67)。因此,我们得出结论,BGS在没有频谱选择性的情况下降低了噪声,并且RETROICOR和BGS对tSNR的改善是相加的。标记持续时间为R-R间期倍数的CASL可用于最小化源自心脏搏动的信号波动。