Differences in displayed pump flow compared to measured flow under varying conditions during simulated cardiopulmonary bypass.

作者信息

Hargrove M, O'Donnell A, Aherne T

机构信息

Cardiothoracic Surgery Unit, Cork University Hospital, Wilton, Cork, Ireland.

出版信息

Perfusion. 2008 Jul;23(4):227-30. doi: 10.1177/0267659108100458.

Abstract

Errors in blood flow delivery due to shunting have been reported to reduce flow by, potentially, up to 40-83% during cardiopulmonary bypass. The standard roller-pump measures revolutions per minute and a calibration factor for different tubing sizes calculates and displays flow accordingly. We compared displayed roller-pump flow with ultrasonically measured flow to ascertain if measured flow correlated with the heart-lung pump flow reading. Comparison of flows was measured under varying conditions of pump run duration, temperature, viscosity, varying arterial/venous loops, occlusiveness, outlet pressure, use of silicone or polyvinyl chloride (PVC) in the roller race, different tubing diameters, and use of a venous vacuum-drainage device.