Hellén-Halme Kristina, Nilsson Mats, Petersson Arne
Department of Oral and Maxillofacial Radiology, Faculty of Odontology, Malmö University, Malmö, Sweden.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 May;107(5):716-20. doi: 10.1016/j.tripleo.2008.12.011. Epub 2009 Feb 6.
The aim of this study was to evaluate if different ways of adjusting brightness and contrast of monitors with different technical standards influence the diagnosis of carious lesions in digital radiographs.
One hundred extracted teeth (premolars and molars) were radiographed. Seven observers evaluated images for approximal carious lesions on 3 monitors: a standard color monitor with brightness and contrast manually adjusted for viewing radiographs, a Digital Imaging and Communication in Medicine (DICOM) part 14 precalibrated color monitor, and a DICOM part 14 precalibrated monochromatic monitor. All evaluations were made in ambient light <50 lux. Receiver operating characteristic curves were plotted to evaluate results. The standard criterion was a histologic examination of sliced teeth. Kappa statistic evaluated intraobserver agreement.
No significant difference in accuracy of approximal carious lesion diagnosis was found between the monitors. Intraobserver agreement varied between fair and good.
No differences that could affect clinicians' abilities to detect carious lesions in digital radiographs existed between the standard monitor and the color and monochrome DICOM part 14 precalibrated monitors.
本研究旨在评估不同技术标准的显示器调整亮度和对比度的不同方式是否会影响数字化X线片上龋损的诊断。
对100颗拔除的牙齿(前磨牙和磨牙)进行X线摄影。7名观察者在3台显示器上评估邻面龋损图像:一台手动调整亮度和对比度以查看X线片的标准彩色显示器、一台医学数字成像和通信(DICOM)第14部分预校准的彩色显示器以及一台DICOM第14部分预校准的单色显示器。所有评估均在环境光<50勒克斯的条件下进行。绘制受试者工作特征曲线以评估结果。标准对照是对切片牙齿进行组织学检查。kappa统计量评估观察者内一致性。
显示器之间在邻面龋损诊断准确性方面未发现显著差异。观察者内一致性在中等至良好之间。
标准显示器与彩色和单色DICOM第14部分预校准显示器之间不存在可能影响临床医生在数字化X线片上检测龋损能力的差异。