Welles Elizabeth G, Hall Amy S, Carpenter D Mark
Department of Pathobiology, Auburn University, Auburn, AL 36849-5519, USA.
Vet Clin Pathol. 2009 Mar;38(1):20-9. doi: 10.1111/j.1939-165X.2008.00084.x. Epub 2008 Oct 28.
With more use of bench-top in-office hematology analyzers, the accuracy of reported values is increasingly important. Instruments use varied methods for cell counting and differentiation, and blood smears may not always be examined.
The purpose of this study was to compare canine CBC results using 4 bench-top instruments (Hemavet 950, Heska CBC-Diff, IDEXX LaserCyte, and IDEXX VetAutoread) with ADVIA 120 and manual leukocyte counts.
EDTA-anticoagulated canine blood samples (n=100) were analyzed on each instrument. Manual differentials were based on 100-cell counts. Linear regression, difference plots, paired t-tests, and estimation of diagnostic equivalence were used to analyze results.
Correlations of HCT, WBC, and platelet counts were very good to excellent between all in-office instruments and the ADVIA 120, but results varied in accuracy (comparability). Hemavet 950 and Heska CBC-Diff results compared best with ADVIA results and manual leukocyte differentials. HCT and platelet counts on the IDEXX VetAutoread compared well with those from the ADVIA. Except for neutrophil counts, leukocyte differentials from all instruments compared poorly with ADVIA and manual counts. Reticulocyte counts on the LaserCyte and VetAutoread compared poorly with those from the ADVIA.
The Hemavet 950 and Heska CBC-Diff performed best of the 4 analyzers we compared. HCT, WBC, and platelet counts on the LaserCyte had minimally sufficient comparability for diagnostic use. Except for neutrophils (granulocytes), leukocyte differential counts were unreliable on all in-office analyzers. Instruments with a 5-part leukocyte differential provided no added benefit over a 3-part differential. Assessment of erythrocyte regeneration on the LaserCyte and VetAutoread was unreliable compared with the ADVIA 120.
随着台式室内血液分析仪的使用越来越多,报告值的准确性变得越来越重要。仪器使用多种细胞计数和分类方法,而且并非总是对血涂片进行检查。
本研究的目的是比较使用4种台式仪器(Hemavet 950、Heska CBC-Diff、IDEXX LaserCyte和IDEXX VetAutoread)得出的犬全血细胞计数(CBC)结果与ADVIA 120及手工白细胞计数结果。
对每份仪器分析100份乙二胺四乙酸(EDTA)抗凝犬血样本。手工分类基于100个细胞计数。采用线性回归、差异图、配对t检验和诊断等效性估计来分析结果。
所有室内仪器与ADVIA 120之间的血细胞比容(HCT)、白细胞(WBC)和血小板计数的相关性非常好至极好,但结果在准确性(可比性)方面存在差异。Hemavet 950和Heska CBC-Diff的结果与ADVIA结果和手工白细胞分类结果相比最佳。IDEXX VetAutoread的HCT和血小板计数与ADVIA的结果比较相符。除中性粒细胞计数外,所有仪器的白细胞分类与ADVIA和手工计数相比差异较大。LaserCyte和VetAutoread的网织红细胞计数与ADVIA的结果相比差异较大。
在我们比较的4种分析仪中,Hemavet 950和Heska CBC-Diff表现最佳。LaserCyte的HCT、WBC和血小板计数在诊断用途上具有最低限度的足够可比性。除中性粒细胞(粒细胞)外,所有室内分析仪的白细胞分类计数都不可靠。具有五分类白细胞分类功能的仪器与三分类相比没有额外优势。与ADVIA 120相比,LaserCyte和VetAutoread对红细胞再生的评估不可靠。