Gurpreet K, Tee G H, Karuthan C
Institute for Public Health, Ministry of Health, Jalan Bangsar, 50590 Kuala Lumpur, Malaysia.
Med J Malaysia. 2008 Aug;63(3):239-43.
A study was undertaken to assess the accuracy of the Omron HEM-907 blood pressure measuring device for use in community studies. A modified version of the British Hypertension Society (BHS) and American Association for the Advancement Medical Instrumentation (AAMI) protocol for measuring the accuracy of a blood pressure measuring device was used. A total of 104 subjects were recruited from two clinics. Observer-observer agreement for readings within the 5 mmHg was good; 80.8% and 84.6% of systolic blood pressure (SBP) and diastolic blood pressure (DBP) agreement respectively. Of the two, the better observer-device agreement readings within the 5 mmHg were 66.4% and 50.0% for SBP and DBP respectively, giving an overall grade B. The mean differences and standard deviation of the differences were within < or =5 mmHg with a standard deviation (SD) of < or =8 mmHg. The Omron HEM-907 satisfied both the AAMI and BHS protocols for accuracy for a non-invasive blood pressure monitoring device using single observer readings.
开展了一项研究,以评估欧姆龙HEM - 907血压测量设备在社区研究中的准确性。使用了英国高血压学会(BHS)和美国医疗仪器促进协会(AAMI)血压测量设备准确性测量方案的修订版。从两家诊所招募了总共104名受试者。观察者之间在5 mmHg范围内读数的一致性良好;收缩压(SBP)和舒张压(DBP)一致性分别为80.8%和84.6%。在这两者中,在5 mmHg范围内观察者与设备更好的一致性读数,SBP和DBP分别为66.4%和50.0%,总体等级为B。差异的均值和标准差在≤5 mmHg范围内,标准差(SD)≤8 mmHg。欧姆龙HEM - 907使用单观察者读数,满足非侵入性血压监测设备准确性的AAMI和BHS方案。