Rapid methods for concurrent measurement of the RF-pulse flip angle and the longitudinal relaxation time.

作者信息

Hsu Jung-Jiin, Zaharchuk Greg, Glover Gary H

机构信息

Richard M. Lucas Center for Imaging, Stanford University, Stanford, California 94305-5488, USA.

出版信息

Magn Reson Med. 2009 Jun;61(6):1319-25. doi: 10.1002/mrm.21900.

Abstract

Measuring both the flip angle (FA) and the longitudinal relaxation time T(1) is essential in quantitative and longitudinal studies because the signal amplitude is dependent on these quantities. Conventional methods can only measure one of them at a time and require long scan times. In this work, two mutually consistent methods are developed; each can acquire multislice data for determining both the FA and T(1) in a scan time about half the time needed for a conventional FA measurement. On the basis of a recent development of longitudinal-relaxation measurement (Hsu and Lowe, J Magn Reson 2004;169:270-278; Hsu and Glover, J Magn Reson 2006;181:98-106), one of the methods uses RF pulse trains of two FAs whereas the other uses pulse trains of different pulse spacing. When only the FA or T(1) is needed, the present methods can still be faster than conventional methods for the needed quantity. In benchmarking with a uniform-density sample, both methods generate precise T(1) values independent of the FA chosen (except at and near 90 degrees ). In the demonstration with three normal volunteers at 3 T, the T(1) values of frontal and occipital white matter, putamen, and caudate are compared; the T(1) values are in agreement with literature values and the intrasubject deviation is 0.2%-2.8%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验