Syringe shape and positioning relative to efficiency volume inside dose calibrators and its role in nuclear medicine quality assurance programs.

作者信息

Santos J A M, Carrasco M F, Lencart J, Bastos A L

机构信息

Serviço de Física Médica, Instituto Português de Oncologia do Porto Francisco Gentil, E.P.E., Rua Dr. António Bernardino de Almeida, 4200072 Porto, Portugal.

出版信息

Appl Radiat Isot. 2009 Jun;67(6):1104-9. doi: 10.1016/j.apradiso.2009.01.084. Epub 2009 Feb 5.

Abstract

A careful analysis of geometry and source positioning influence in the activity measurement outcome of a nuclear medicine dose calibrator is presented for (99m)Tc. The implementation of a quasi-point source apparent activity curve measurement is proposed for an accurate correction of the activity inside several syringes, and compared with a theoretical geometric efficiency model. Additionally, new geometrical parameters are proposed to test and verify the correct positioning of the syringes as part of acceptance testing and quality control procedures.

摘要

针对(99m)Tc,本文对核医学剂量校准器活度测量结果中的几何形状和源定位影响进行了细致分析。为准确校正多个注射器内的活度,提出了准点源表观活度曲线测量方法,并与理论几何效率模型进行了比较。此外,还提出了新的几何参数,用于测试和验证注射器的正确定位,作为验收测试和质量控制程序的一部分。