Hoang Dinh Xuan, Vu Truong Dinh, Dan Do Van, Nhan Nguyen Thanh
Center for Radiation Protection, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.
Center for Research and Production of Radioisotopes, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.
Indian J Nucl Med. 2024 Jul-Aug;39(4):272-278. doi: 10.4103/ijnm.ijnm_51_24. Epub 2024 Nov 18.
This study aimed to evaluate the principal technical characteristics of a well-type gas-filled ionization chamber dose calibrator used in measuring radiopharmaceutical activity, namely accuracy, repeatability, and linearity. Furthermore, this work also explored the correlation between the device's response and the position and volume of the radiopharmaceutical I-131.
Experimental measurements were conducted on the ATOMLAB 500 dose calibrator using NIST traceable Cs-137 source to determine the accuracy and repeatability. For the linearity test, the Tc-99m solution produced at the Dalat Nuclear Research Institute, Vietnam was utilized to examine the performance of the device over a wide measurement range. Effects resulting from different volumes and measuring positions were also determined by experiments and Monte Carlo simulations.
Based on acceptance test results, it is revealed that the dose calibrator's characteristics comply with international standards, where the deviations of accuracy, repeatability, and linearity are all lower than 1.0%. The response of the dose calibrator to different measuring positions and volumes was well controlled. To rectify the discrepancy in the response to the changes in solution volume, correction functions were proposed.
It is concluded that the dose calibrator is suitable for radioactivity quantification and adheres to standards recommended by international organizations for nuclear medicine activities. The outcome of this work will serve as the foundation for establishing a standard for I-131 radiopharmaceuticals' radioactivity in Vietnam.
本研究旨在评估用于测量放射性药物活度的井型充气电离室剂量校准仪的主要技术特性,即准确性、重复性和线性。此外,本研究还探讨了该设备的响应与放射性药物I-131的位置和体积之间的相关性。
使用可溯源至美国国家标准与技术研究院(NIST)的Cs-137源对ATOMLAB 500剂量校准仪进行实验测量,以确定其准确性和重复性。对于线性测试,使用越南大叻核研究所生产的Tc-99m溶液来检验该设备在宽测量范围内的性能。还通过实验和蒙特卡洛模拟确定了不同体积和测量位置所产生的影响。
根据验收测试结果,表明剂量校准仪的特性符合国际标准,准确性、重复性和线性的偏差均低于1.0%。剂量校准仪对不同测量位置和体积的响应得到了很好的控制。为纠正对溶液体积变化响应中的差异,提出了校正函数。
得出结论,该剂量校准仪适用于放射性定量,符合国际组织推荐的核医学活动标准。本研究结果将作为越南建立I-131放射性药物放射性标准的基础。