Ly Suw Young, Kim Min Joon
Biosensor Research Institute, Seoul National University of Technology, 172 gongreung 2-dong, Nowon-gu, Seoul, South Korea.
J Clin Lab Anal. 2009;23(2):82-7. doi: 10.1002/jcla.20294.
A voltammetric diagnosis of a chromium (VI) ion was investigated using a fluorine-doped graphite pencil electrode. Square wave (SW) stripping working conditions were attained at a high range of 0.051-0.45 mg L(-1) and a microrange of 0.05-0.4 microg L(-1) in a 0.1 M NH(4)H(2)PO(4) electrolyte solution, at a relative standard deviation of 1.68% (RSD, n=15), using 10.0 microg L(-1) Cr(VI). A fast experimental time was used only for the 120 sec SW accumulation time. An analytical detection-limit (DL) of 0.008 microg L(-1) was attained. DL appeared to be more sensitive than common voltammetric and spectrophotometric assays. The developed sensor was applied to tap water and the urine of a smoker. It was found that the methods can be applicable for in vivo fluid or medicinal diagnosis.
使用氟掺杂石墨铅笔电极对六价铬离子进行了伏安诊断。在0.1 M NH₄H₂PO₄电解质溶液中,采用方波(SW)溶出法,在0.051 - 0.45 mg L⁻¹的高浓度范围和0.05 - 0.4 μg L⁻¹的微浓度范围内,以10.0 μg L⁻¹的Cr(VI)为标准,相对标准偏差为1.68%(RSD,n = 15)。快速实验时间仅用于120秒的SW富集时间。获得了0.008 μg L⁻¹的分析检测限(DL)。DL似乎比普通伏安法和分光光度法更灵敏。所开发的传感器应用于自来水和吸烟者的尿液。结果发现该方法可用于体内液体或药物诊断。