Louie Richard F, Sumner Stephanie L, Belcher Shaunyé, Mathew Ron, Tran Nam K, Kost Gerald J
University of California, Davis-Lawrence Livermore National Laboratory Point-of-Care Technologies Center, CA, USA.
Disaster Med Public Health Prep. 2009 Mar;3(1):13-7. doi: 10.1097/DMP.0b013e3181979a06.
Point-of-care testing (POCT) devices are deployed in the field for emergency on-site testing under a wide range of environmental conditions. Our objective was to evaluate the performance of glucose meter test strips and handheld blood gas analyzer cartridges following thermal stresses that simulate field conditions.
We evaluated electrochemical and spectrophotometric glucose meter systems and a handheld blood gas analyzer. Glucose test strips were cold-stressed (-21 degrees C) and heat-stressed (40 degrees C) for up to 4 weeks. Blood gas cartridges were stressed at -21 degrees C, 2 degrees C, and 40 degrees C for up to 72 hours. Test strip and cartridge performance was evaluated using aqueous quality control solutions. Results were compared with those obtained with unstressed POCT strips and cartridges.
Heated glucose test strips and blood gas cartridges yielded elevated results. Frozen test strips and cooled cartridges yielded depressed glucose and blood gas results, respectively. Frozen cartridges failed.
The performance of glucose test strips and blood gas cartridges was affected adversely by thermal stresses. Heating generated elevated results, and cooling depressed results. Disaster medical assistance teams and emergency medical responders should be aware of these risks. Field POCT devices must be robust to withstand adverse conditions. We recommend that industry produce POCT devices and reagents suitable for disaster medical assistance teams.
即时检验(POCT)设备部署在现场,用于在各种环境条件下进行紧急现场检测。我们的目的是评估血糖仪测试条和手持式血气分析仪试剂盒在模拟现场条件的热应激后的性能。
我们评估了电化学和分光光度血糖仪系统以及手持式血气分析仪。葡萄糖测试条在-21℃下进行冷应激和在40℃下进行热应激,持续长达4周。血气试剂盒在-21℃、2℃和40℃下进行应激处理,持续长达72小时。使用水性质量控制溶液评估测试条和试剂盒的性能。将结果与未受应激的POCT测试条和试剂盒所获得的结果进行比较。
加热后的葡萄糖测试条和血气试剂盒产生的结果升高。冷冻的测试条和冷却的试剂盒分别产生降低的葡萄糖和血气结果。冷冻的试剂盒失效。
葡萄糖测试条和血气试剂盒的性能受到热应激的不利影响。加热产生升高的结果,冷却产生降低的结果。灾难医疗援助团队和紧急医疗救援人员应意识到这些风险。现场POCT设备必须坚固耐用以承受不利条件。我们建议行业生产适合灾难医疗援助团队的POCT设备和试剂。