Abstract

The present study compares two methods of haemoglobin measurement during surgery in 40 patients aged 1 to 85 years. The reference method is the Coulter STKS based on red cell count and mean cellular content in haemoglobin measurement. The method tested (Hemocue-Diagnostics Transfusion) allows instantaneous determination of haemoglobin on a 10 mu sample by measuring infrared absorption by haemoglobin in the 565 and 880 nm wavelengths. These methods have been compared using the method of assessment of agreement described by Bland and Altman. As previous reports suggested that the precision of haemoglobin measurement by Hemocue could depend on the capillary or venous site of sampling, we assessed agreement of the two methods of measurement of haemoglobin at both sites. In addition, the relationship between hematocrit and haemoglobin measured by either method has been assessed. On venous samples, there was a good agreement between the two methods of measurement of haemoglobin (n = 47), with a mean difference of -0.166 g/dl-1 and agreement limits of -1.041 g/dl-1 and +0.195 g/dl-1. Ninety six percent of the observed differences to two methods were within the 95% confidence interval. On capillary samples, there was a poor agreement between Hemocue and Coulter since the mean difference was -0.272 g/dl-1 with the following limits of agreement: -2.356 g/dl-1 and +1.812 g/dl-1. Only 92% of the observed differences were within the 95% confidence interval, with 17% of these being greater than the clinically accepted limit of 1 g/dl-1. The precision of Hemocue measurement of capillary blood haemoglobin was only +/- 16%.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

本研究比较了40名年龄在1至85岁的患者在手术期间两种血红蛋白测量方法。参考方法是基于红细胞计数和血红蛋白测量中平均细胞含量的库尔特STKS。所测试的方法(Hemocue诊断输血法)通过测量血红蛋白在565和880纳米波长处的红外吸收,能够在10微升样本上即时测定血红蛋白。使用布兰德和奥特曼描述的一致性评估方法对这些方法进行了比较。正如先前的报告所表明的,Hemocue测量血红蛋白的精度可能取决于采样的毛细血管或静脉部位,我们评估了在这两个部位两种血红蛋白测量方法的一致性。此外,还评估了两种方法所测血细胞比容与血红蛋白之间的关系。在静脉样本上,两种血红蛋白测量方法之间具有良好的一致性(n = 47),平均差异为-0.166克/分升,一致性界限为-1.041克/分升和+0.195克/分升。两种方法观察到的差异中有96%在95%置信区间内。在毛细血管样本上,Hemocue和库尔特之间的一致性较差,因为平均差异为-0.272克/分升,一致性界限如下:-2.356克/分升和+1.812克/分升。只有92%的观察到的差异在95%置信区间内,其中17%大于临床可接受的1克/分升的界限。Hemocue测量毛细血管血血红蛋白的精度仅为+/- 16%。(摘要截断于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验