Suppr超能文献

准分子原位角膜磨镶术后非接触式眼压测量的可靠性

Reliability of noncontact pachymetry after laser in situ keratomileusis.

作者信息

Maldonado Miguel J, López-Miguel Alberto, Nieto Juan C, Cano-Parra Juan, Calvo Begoña, Alió Jorge L

机构信息

Department of Ophthalmology, University Clinic, University of Navarra, Pamplona, Spain.

出版信息

Invest Ophthalmol Vis Sci. 2009 Sep;50(9):4135-41. doi: 10.1167/iovs.09-3408. Epub 2009 Apr 8.

Abstract

PURPOSE

To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements obtained by combined scanning-slit/Placido-disc topography (Orbscan II; Bausch & Lomb, Rochester, NY) after laser in situ keratomileusis (LASIK) and to compare the results with another noncontact method, specular microscopy (SP-2000P; Topcon, Tokyo, Japan).

METHODS

To analyze intrasession repeatability, one examiner measured 22 postmyopic LASIK eyes 10 times successively in the shortest time possible, using both devices randomly. To study intersession reproducibility, the same operator obtained measurements from another 50 eyes with stable refraction in two consecutive visits at the same time of the day between 6 and 9 months after myopic LASIK. Any association between residual stromal bed thickness and measurement variability was recorded and evaluated.

RESULTS

For intrasession repeatability, Orbscan II and Topcon SP-2000P CCT measurements showed a repeatability of 20.2 (4.3%) and 12.8 (2.6%) microm, respectively. Both devices yielded excellent intraclass correlation coefficients (ICC; 0.98 and 0.99, respectively). For intersession reproducibility, no difference in CCT measurements was found. The coefficient of intersession reproducibility were 6.7% (29.5 microm) for Orbscan II and 4.3% (19.5 microm) for Topcon SP-2000P CCT measurements. The ICCs (0.95 and 0.96, respectively) indicated good intersession reliability. Repeatability and reproducibility with both devices were unrelated to stromal bed thickness.

CONCLUSIONS

Both noncontact pachymeters provide repeatable CCT measurements in transparent postmyopic LASIK corneas after the early postoperative period. Intersession variations in CCT of more than 29 microm with the Orbscan II and 19 microm with the Topcon SP-2000P may reflect true corneal change. These estimates should help investigators and clinicians differentiate actual CCT modification from measurement variability.

摘要

目的

评估准分子原位角膜磨镶术(LASIK)后使用联合扫描裂隙/普拉西多盘角膜地形图仪(Orbscan II;博士伦公司,纽约州罗切斯特市)测量中央角膜厚度(CCT)的可重复性和再现性,并将结果与另一种非接触式方法——镜面显微镜检查(SP - 2000P;拓普康公司,日本东京)进行比较。

方法

为分析测量过程中的可重复性,一名检查者在尽可能短的时间内,使用这两种设备随机对22只近视LASIK术后眼依次测量10次。为研究不同测量时段间的再现性,同一名操作者在近视LASIK术后6至9个月的同一天同一时间,对另外50只屈光稳定的眼睛进行连续两次检查并测量。记录并评估剩余基质床厚度与测量变异性之间的任何关联。

结果

对于测量过程中的可重复性,Orbscan II和拓普康SP - 2000P测量的CCT可重复性分别为20.2(4.3%)和12.8(2.6%)微米。两种设备均产生了出色的组内相关系数(ICC;分别为0.98和0.99)。对于不同测量时段间的再现性,CCT测量未发现差异。Orbscan II测量的不同测量时段间再现性系数为6.7%(29.5微米),拓普康SP - 2000P测量的CCT为4.3%(19.5微米)。ICC(分别为0.95和0.96)表明不同测量时段间可靠性良好。两种设备的可重复性和再现性均与基质床厚度无关。

结论

两种非接触式测厚仪在术后早期后能为透明的近视LASIK术后角膜提供可重复的CCT测量。使用Orbscan II时CCT在不同测量时段间变化超过29微米,使用拓普康SP - 2000P时超过19微米,可能反映了角膜的真实变化。这些评估结果应有助于研究人员和临床医生区分实际的CCT改变与测量变异性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验