使用角膜动态Scheimpflug分析仪Corvis ST评估中央角膜厚度,并与Pentacam旋转Scheimpflug系统和超声测厚法在正常眼中进行比较。

Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

作者信息

Yu Ayong, Zhao Weiqi, Savini Giacomo, Huang Zixu, Bao Fangjun, Lu Weicong, Wang Qinmei, Huang Jinhai

机构信息

School of Ophthalmology and Optometry, Wenzhou Medical University, 270 West Xueyuan Road, Wenzhou, Zhejiang 325027, China ; Key Laboratory of Vision Science, Ministry of Health of the People's Republic of China, Wenzhou, Zhejiang, China.

G. B. Bietti Foundation IRCCS, Rome, Italy.

出版信息

J Ophthalmol. 2015;2015:767012. doi: 10.1155/2015/767012. Epub 2015 Dec 1.

Abstract

Purpose. To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements by corneal dynamic Scheimpflug analyzer Corvis ST in normal eyes and compare the agreement with Pentacam rotating Scheimpflug System and ultrasound pachymetry. Methods. 84 right eyes underwent Corvis ST measurements performed by two operators. The test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to evaluate the intraoperator repeatability and interoperator reproducibility. CCT measurements also were obtained from Pentacam and ultrasound pachymetry by the first operator. The agreement between the three devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots. Results. Corvis ST showed high repeatability as indicated by TRT ≤ 13.0 μm, CoV < 0.9%, and ICC > 0.97. The interoperator reproducibility was also excellent. The CoV was <0.9%, and ICC was >0.97. Corvis ST showed significantly lower values than Pentacam and ultrasound pachymetry (P < 0.001). The 95% LoA between Corvis ST and Pentacam or ultrasound pachymetry were -15.8 to 9.5 μm and -27.9 to 12.3 μm, respectively. Conclusions. Corvis ST showed excellent repeatability and interoperator reproducibility of CCT measurements in normal eyes. Corvis ST is interchangeable with Pentacam but not with ultrasound pachymetry.

摘要

目的。评估角膜动态Scheimpflug分析仪Corvis ST测量正常眼中央角膜厚度(CCT)的可重复性和再现性,并与Pentacam旋转Scheimpflug系统和超声测厚法进行一致性比较。方法。84只右眼由两名操作人员进行Corvis ST测量。采用重测重复性(TRT)、受试者内变异系数(CoV)和组内相关系数(ICC)评估操作人员内重复性和操作人员间再现性。第一名操作人员还通过Pentacam和超声测厚法获得了CCT测量值。用95%一致性界限(LoA)和Bland-Altman图评估三种设备之间的一致性。结果。Corvis ST显示出高重复性,TRT≤13.0μm、CoV<0.9%和ICC>0.97表明了这一点。操作人员间再现性也非常好。CoV<0.9%,ICC>0.97。Corvis ST显示的值明显低于Pentacam和超声测厚法(P<0.001)。Corvis ST与Pentacam或超声测厚法之间的95%LoA分别为-15.8至9.5μm和-27.9至12.3μm。结论。Corvis ST在正常眼中显示出CCT测量的优异重复性和操作人员间再现性。Corvis ST可与Pentacam互换,但不能与超声测厚法互换。