Morales-Cid Gabriel, Cárdenas Soledad, Simonet Bartolomé M, Valcárcel Miguel
Department of Analytical Chemistry, Campus de Rabanales, University of Cordoba, Cordoba, Spain.
Electrophoresis. 2009 May;30(10):1684-91. doi: 10.1002/elps.200800493.
This paper reports for the first time the use of microextraction by packed sorbent in combination with CE. The combined system was used to determine anesthetic drugs in human plasma. A microdialysis fiber was coupled on-line to the microextraction unit in order to distinguish between free and total concentrations of drugs. The system was automated by connecting the microextraction unit to a syringe pump and interfacing it to a computer. The ensuing method allows the determination of 10 microg/L concentrations of free drugs and 1 microg/L concentrations of total drugs from only 200 microL of sample with an RSD of less than 9%.