Suppr超能文献

使用远程探头麦克风测量来验证助听器性能。

Remote probe microphone measurement to verify hearing aid performance.

作者信息

Ferrari Deborah Viviane, Bernardez-Braga Gabriela Rosito Alvarez

机构信息

Department of Speech Language Pathology and Audiology, Dentistry School of Bauru, University of São Paulo, Bauru, SP, Brazil.

出版信息

J Telemed Telecare. 2009;15(3):122-4. doi: 10.1258/jtt.2009.003005.

Abstract

We assessed the feasibility of obtaining probe microphone measurements of hearing aids at a distance. Face-to-face and remote probe microphone measurements were carried out in 60 hearing aid users (mean age 67 yrs) with uni- or bilateral hearing losses (105 ears tested). The participant and a facilitator were located in a room equipped with a probe microphone system interfaced to a PC. Desktop videoconferencing and application sharing was used to allow an audiologist in another room to instruct the facilitator and control the equipment via the LAN. There were significant correlations between face-to-face and remote real ear unaided response (REUR), aided response (REAR) and insertion gain (REIG) at seven discrete frequencies from 250 to 6000 Hz. Differences between face-to-face and remote responses were within the reported variability for probe microphone measurements themselves. The results show that remote probe microphone measurements are feasible and might improve the quality of public hearing aid services and professional training in Brazil.

摘要

我们评估了在一定距离获取助听器探头麦克风测量值的可行性。对60名患有单侧或双侧听力损失的助听器使用者(平均年龄67岁,共测试105只耳朵)进行了面对面和远程探头麦克风测量。参与者和一名协助者位于配备有与电脑相连的探头麦克风系统的房间内。使用桌面视频会议和应用程序共享,以便另一个房间的听力学家通过局域网指导协助者并控制设备。在250至6000赫兹的七个离散频率上,面对面和远程真耳未助听反应(REUR)、助听反应(REAR)和插入增益(REIG)之间存在显著相关性。面对面和远程反应之间的差异在探头麦克风测量本身所报告的变异性范围内。结果表明,远程探头麦克风测量是可行的,可能会提高巴西公共助听器服务和专业培训的质量。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验