Purton David G, Chandler Nicholas P, Monteith Brian D, Qualtrough Alison J
Eur J Prosthodont Restor Dent. 2009 Mar;17(1):30-4.
The aim was to determine whether extracted teeth could be used to test the Prepometer instrument, which indicates pulp proximity with green/amber/red light emitting diodes. Third molars were reduced to a plane in dentine and Prepometer readings made. Dentine was removed in 0.5 mm increments and readings made until only lights 9 or 10 (red) showed. The teeth were sectioned and the dentine thickness measured. Analysis permitted construction of a highly significant predictor-model (p < 0.01), the red/amber light boundary coinciding with a dentine thickness of 2.4 mm. The Prepometer was consistent in predicting pulp proximity but was more sensitive than specified.