在进行内镜逆行胰胆管造影(ERCP)时,比较固态导管和水灌注三腔导管用于Oddi括约肌测压的前瞻性试验。

Prospective trial comparing solid-state catheter and water-perfusion triple-lumen catheter for sphincter of Oddi manometry done at the time of ERCP.

作者信息

Draganov Peter V, Kowalczyk Lukasz, Forsmark Chris E

机构信息

Division of Gastroenterology, Hepatology and Nutrition, University of Florida, Gainesville, Florida 32610, USA.

出版信息

Gastrointest Endosc. 2009 Jul;70(1):92-5. doi: 10.1016/j.gie.2008.11.044. Epub 2009 Apr 21.

Abstract

BACKGROUND

Sphincter of Oddi manometry is the reference standard for the diagnosis of sphincter of Oddi dysfunction. Numerous studies have established ranges of normal values as well as typical readings in pathologic conditions. All these studies have been done using a water-perfused, triple-lumen catheter (TLC). A recently approved, new generation, solid-state catheter (SSC) has potential advantages, but concerns have been raised as to whether the pressures obtained by TLCs are reproducible by SSCs. To date, no data exist on the accuracy of sphincter of Oddi pressure measurements with the new-generation SSCs.

OBJECTIVE

To evaluate the accuracy of the SSC by using the TLC as the reference standard.

DESIGN

Prospective crossover trial.

SETTING

A tertiary academic center.

PATIENTS

Thirty patients with clinical indications for sphincter of Oddi manometry.

INTERVENTIONS

Sphincter of Oddi manometry with TLC and SSC in the same patient.

MAIN OUTCOME MEASUREMENTS

Accuracy of sphincter of Oddi pressure measurements.

RESULTS

A total of 376 pressure measurements in 47 sphincter segments (24 biliary, 23 pancreatic) were obtained. Manometry results were abnormal in 10 of 24 biliary sphincters and 12 of 23 pancreatic sphincters. There was complete agreement on the final results of the sphincter of Oddi manometry (normal/abnormal) between the TLC and SSC (accuracy 100%). A split-plot analysis of the 378 individual measurements was performed. The P value of .9966 was insignificant, consistent with no catheter effect on the measurements.

LIMITATIONS

Lack of blinding.

CONCLUSIONS

Measurement of sphincter of Oddi pressures with the SSC is accurate, and results were essentially identical to those of the water-perfused catheter system.

摘要

背景

Oddi括约肌测压是诊断Oddi括约肌功能障碍的参考标准。众多研究已确定了正常数值范围以及病理情况下的典型读数。所有这些研究均使用水灌注三腔导管(TLC)完成。最近获批的新一代固态导管(SSC)具有潜在优势,但有人担心TLC所测压力能否由SSC重现。迄今为止,尚无关于新一代SSC测量Oddi括约肌压力准确性的数据。

目的

以TLC作为参考标准评估SSC的准确性。

设计

前瞻性交叉试验。

地点

一家三级学术中心。

患者

30例有Oddi括约肌测压临床指征的患者。

干预措施

对同一患者使用TLC和SSC进行Oddi括约肌测压。

主要观察指标

Oddi括约肌压力测量的准确性。

结果

在47个括约肌节段(24个胆管、23个胰管)共获得376次压力测量结果。24个胆管括约肌中有10个测压结果异常,23个胰管括约肌中有12个异常。TLC和SSC对Oddi括约肌测压的最终结果(正常/异常)完全一致(准确性100%)。对378次个体测量结果进行了裂区分析。P值为0.9966,无统计学意义,表明导管对测量结果无影响。

局限性

缺乏盲法。

结论

使用SSC测量Oddi括约肌压力是准确的,结果与水灌注导管系统基本相同。