Kawamoto Masahiko, Geenen Joseph, Omari Taher, Schloithe Ann C, Saccone Gino T P, Toouli James
Department of General and Digestive Surgery, Flinders University, Flinders Medical Centre, Bedford Park, SA 5042, Australia.
J Hepatobiliary Pancreat Surg. 2008;15(4):391-6. doi: 10.1007/s00534-007-1262-1. Epub 2008 Aug 1.
BACKGROUND/PURPOSE: Perfused multilumen sphincter of Oddi (SO) manometry is accepted as the gold standard for diagnosis of SO dysfunction. However, this technique is associated with a relatively high incidence of post-procedure acute pancreatitis. In addition, triple-lumen manometry recordings may be difficult to interpret, as movement may produce artifacts. We have refined the development of a sleeve sensor for human SO manometry. This assembly aims to overcome the above limitations. In this study the accuracy of sleeve SO manometry (SOM) has been evaluated and compared with standard triple-lumen perfused SOM.
Patients undergoing SO manometric studies consented to having both standard triple-lumen and sleeve SOM. A total of 32 paired studies were performed in 29 patients. Diagnosis was made only from standard triple-lumen SOM and the patient treated accordingly. For each study, SO basal pressure, contraction, amplitude, and frequency were recorded.
There was no statistically significant difference in the recordings of SO basal pressure, contraction, amplitude, and frequency between the two techniques. A strong correlation was demonstrated between SO basal pressure determined with the two catheters. The accuracy of sleeve SOM is comparable to standard triple-lumen SOM, with less movement artifact. One patient developed mild post-manometric pancreatitis.
The sleeve catheter records SO pressures with comparable values to standard triple-lumen SOM. The sleeve assembly potentially can replace the use of the perfused triple-lumen catheter for the objective diagnosis of SO dysfunction.
背景/目的:Oddi括约肌(SO)灌注式多腔测压法被公认为诊断SO功能障碍的金标准。然而,该技术与术后急性胰腺炎的相对高发生率相关。此外,三腔测压记录可能难以解读,因为移动可能产生伪影。我们改进了用于人体SO测压的套管传感器的开发。该组件旨在克服上述局限性。在本研究中,评估了套管SO测压法(SOM)的准确性,并与标准三腔灌注式SOM进行了比较。
接受SO测压研究的患者同意同时进行标准三腔和套管SOM。在29例患者中总共进行了32对研究。仅根据标准三腔SOM做出诊断并相应地对患者进行治疗。对于每项研究,记录SO基础压力、收缩、幅度和频率。
两种技术在SO基础压力、收缩、幅度和频率的记录方面无统计学显著差异。用两根导管测定的SO基础压力之间显示出强相关性。套管SOM的准确性与标准三腔SOM相当,且移动伪影较少。1例患者发生了轻度测压后胰腺炎。
套管导管记录的SO压力值与标准三腔SOM相当。套管组件有可能取代灌注式三腔导管用于SO功能障碍的客观诊断。