Fernández-Vega Luis, Madrid-Costa David, Alfonso José F, Montés-Micó Robert, Poo-López Arancha
Fernández-Vega Ophthalmological Institute, Surgery Department, School of Medicine, University of Oviedo, Oviedo, Spain.
J Cataract Refract Surg. 2009 May;35(5):825-32. doi: 10.1016/j.jcrs.2008.12.040.
To evaluate the optical and visual quality in eyes with diffractive intraocular lenses (IOLs) after myopic laser in situ keratomileusis (LASIK).
Fernández-Vega Ophthalmological Institute, Oviedo, Spain.
The best corrected visual acuity (BCVA [distance]) at 12.5%, 25.0%, and 100% contrasts under photopic and mesopic conditions; best distance-corrected near acuity; corneal total higher-order aberrations (HOAs); and optical quality by modulation transfer function (MTF) were measured after myopic LASIK in eyes with a spherical AcrySof ReSTOR SN60D3 IOL or an aspheric Acri.LISA 366D IOL and phakic eyes (control).
Under photopic conditions at 100% contrast, the BCVA was 0.1 logMAR or better in all groups. At 25.0% and 12.5% contrasts, there were no statistically significant differences between the aspheric IOL group and the spherical IOL or control group (P >.01). The control group had better BCVA than the spherical IOL group (P = .0002, 25.0%; P<.0001, 12.5%). Under mesopic conditions, the aspheric IOL group had better BCVA than the spherical IOL group at all contrasts (P<.01). There was no statistically significant difference between the 2 IOL groups in mean best distance-corrected near acuity or between all 3 groups in HOAs (P >.01). With 6.0 mm and 3.0 mm pupils, there were no statistically significant differences in MTF between the aspheric IOL group and the control group; both groups had better MTF than the spherical IOL group with a 6.0 mm pupil (P<.01).
After LASIK, the aspheric IOL provided better visual quality and optical quality than spherical IOL under mesopic conditions (large pupil).
评估近视激光原位角膜磨镶术(LASIK)后植入衍射型人工晶状体(IOL)的眼睛的光学和视觉质量。
西班牙奥维耶多的费尔南德斯 - 维加眼科研究所。
测量了植入球面AcrySof ReSTOR SN60D3 IOL或非球面Acri.LISA 366D IOL的眼睛以及有晶状体眼(对照组)在近视LASIK术后,在明视和微光条件下12.5%、25.0%和100%对比度时的最佳矫正视力(BCVA[远距离]);最佳远距离矫正近视力;角膜总高阶像差(HOA);以及通过调制传递函数(MTF)测量的光学质量。
在明视条件下100%对比度时,所有组的BCVA均为0.1 logMAR或更好。在25.0%和12.5%对比度时,非球面IOL组与球面IOL组或对照组之间无统计学显著差异(P>.01)。对照组的BCVA优于球面IOL组(P =.0002,25.0%;P<.0001,12.5%)。在微光条件下,非球面IOL组在所有对比度下的BCVA均优于球面IOL组(P<.01)。两个IOL组在平均最佳远距离矫正近视力方面无统计学显著差异,所有三组在HOA方面也无统计学显著差异(P>.01)。对于6.0 mm和3.0 mm的瞳孔,非球面IOL组与对照组之间的MTF无统计学显著差异;两组在6.0 mm瞳孔时的MTF均优于球面IOL组(P<.01)。
LASIK术后,在微光条件下(大瞳孔),非球面IOL比球面IOL提供更好的视觉质量和光学质量。