Abstract

The reagent carrier Reflotron-Creatinine was tested for its suitability for whole blood and plasma samples of dogs and cats. It showed a very good correlation between the results of Reflotron-Creatinine and the reference method. Using plasma the results of both methods were extremely close. Using whole blood with a hematocrit of up to 50% good results could be achieved. If the hematocrit exceeded 50% large deviations could result.

摘要

对试剂载体Reflotron - 肌酐进行了测试,以确定其对犬猫全血和血浆样本的适用性。Reflotron - 肌酐的结果与参考方法之间显示出非常良好的相关性。使用血浆时,两种方法的结果极为接近。使用血细胞比容高达50%的全血可获得良好结果。如果血细胞比容超过50%,可能会导致较大偏差。