Kridde G, Hirschberger J
I. Medizinischen Tierklinik, Ludwig-Maximilians-Universität München.
Tierarztl Prax. 1991 Aug;19(4):447-9.
The reagent carrier Reflotron-Creatinine was tested for its suitability for whole blood and plasma samples of dogs and cats. It showed a very good correlation between the results of Reflotron-Creatinine and the reference method. Using plasma the results of both methods were extremely close. Using whole blood with a hematocrit of up to 50% good results could be achieved. If the hematocrit exceeded 50% large deviations could result.