Suppr超能文献

带有微螺纹的粗糙表面种植体颈部可减少边缘骨吸收量:一项前瞻性临床研究。

A rough surface implant neck with microthreads reduces the amount of marginal bone loss: a prospective clinical study.

作者信息

Bratu Emanuel A, Tandlich Moshik, Shapira Lior

机构信息

Department of Implant Restorations, University of Medicine and Pharmacy, Timisoara School of Dental Medicine, Timisoara, Romania. [corrected]

出版信息

Clin Oral Implants Res. 2009 Aug;20(8):827-32. doi: 10.1111/j.1600-0501.2009.01730.x. Epub 2009 Jun 7.

Abstract

OBJECTIVES

An intra-individual controlled clinical trial was conducted to evaluate and compare the amount of marginal bone loss (MBL) found around implants of a comparable design, with or without retention grooves (microthreads) or polished necks, during the early stages of healing.

MATERIALS AND METHODS

Forty-eight (48) patients with missing mandibular posterior teeth were treated with two commercially available implants of the same brand (MIS): one with microthreads (S-model) and the other with a polished neck (L-model). MBL around each implant was measured on follow-up radiograms taken 4 months after placement (exposure and crown cementation), and 6 and 12 months after loading.

RESULTS

Forty-six (46) patients completed the study, making 46 implant pairs available for statistical analysis. None of the implants failed to integrate. All the implants displayed some extent of bone loss throughout the follow-up period. At each time point (exposure, 6 and 12 months after loading), the S-model implants displayed statistically significant lower amounts of bone loss (0.22 vs. 0.76, 0.57 vs. 1.22 and 0.9 vs. 1.5 mm, respectively). Other than the type of the implant, no correlation was found between MBL and the implant stability values (PerioTest), dimensions, site of insertion or any of the other collected variables.

CONCLUSIONS

Implants with a roughened neck surface and microthreads are more resistant to MBL during the first phases of healing, as compared with implants with a polished neck.

摘要

目的

进行一项个体内对照临床试验,以评估和比较在愈合早期,具有或不具有固位槽(微螺纹)或抛光颈部的类似设计种植体周围的边缘骨丢失(MBL)量。

材料与方法

48例下颌后牙缺失患者接受了同一品牌(MIS)的两种市售种植体治疗:一种带有微螺纹(S型),另一种带有抛光颈部(L型)。在种植体植入后4个月(暴露和冠修复粘结)、加载后6个月和12个月拍摄的随访X光片上测量每个种植体周围的MBL。

结果

46例患者完成了研究,共46对种植体可供统计分析。所有种植体均成功整合。在整个随访期间,所有种植体均出现了一定程度的骨丢失。在每个时间点(暴露、加载后6个月和12个月),S型种植体的骨丢失量在统计学上显著较低(分别为0.22 vs. 0.76、0.57 vs. 1.22和0.9 vs. 1.5 mm)。除种植体类型外,未发现MBL与种植体稳定性值(牙周探测)、尺寸、植入部位或任何其他收集的变量之间存在相关性。

结论

与颈部抛光的种植体相比,颈部表面粗糙且带有微螺纹的种植体在愈合的第一阶段对MBL更具抵抗力。