Xie Tuqiang, Guo Shuguang, Chen Zhongping, Mukai David, Brenner Matthew
Opt Express. 2006 Apr 17;14(8):3238-46. doi: 10.1364/oe.14.003238.
In this manuscript, a GRIN (gradient index) lens rod based probe for endoscopic spectral domain optical coherence tomography (OCT) with dynamic focus tracking is presented. Current endoscopic OCT systems have a fixed focal plane or working distance. In contrast, the focus of this endoscopic OCT probe can dynamically be adjusted at a high speed (500 mm/s) without changing reference arm length to obtain high quality OCT images for contact or non-contact tissue applications, or for areas of difficult access for probes. The dynamic focusing range of the probe can be from 0 to 7.5 mm without moving the probe itself. The imaging depth is 2.8 mm and the lateral scanning range is up to 2.7 mm or 4.5 mm (determined by the diameter of different GRIN lens rods). Three dimensional imaging can be performed using this system over an area of tissue corresponding to the GRIN lens surface. The experimental results demonstrate that this GRIN lens rod based OCT system can perform a high quality non-contact in vivo imaging. This rigid OCT probe is solid and can be adapted to safely access internal organs, to perform front or side view imaging with an imaging speed of 8 frames per second, with all moving parts proximal to the GRIN lens, and has great potential for use in extremely compact OCT endoscopes for in vivo imaging in both biological research and clinical applications.
在本论文中,提出了一种基于梯度折射率(GRIN)透镜棒的探头,用于具有动态聚焦跟踪功能的内窥光谱域光学相干断层扫描(OCT)。当前的内窥OCT系统具有固定的焦平面或工作距离。相比之下,这种内窥OCT探头的焦点可以在不改变参考臂长度的情况下以高速(500毫米/秒)动态调整,以获取用于接触或非接触组织应用的高质量OCT图像,或用于探头难以到达的区域。探头的动态聚焦范围可以在不移动探头本身的情况下从0到7.5毫米。成像深度为2.8毫米,横向扫描范围可达2.7毫米或4.5毫米(由不同GRIN透镜棒的直径决定)。使用该系统可以在与GRIN透镜表面相对应的组织区域上进行三维成像。实验结果表明,这种基于GRIN透镜棒的OCT系统可以进行高质量的非接触式体内成像。这种刚性OCT探头坚固耐用,可安全地用于进入内部器官,以每秒8帧的成像速度进行前视或侧视成像,所有运动部件都靠近GRIN透镜,在用于生物研究和临床应用的体内成像的极其紧凑的OCT内窥镜中具有巨大的应用潜力。