基于干燥人类头骨的锥形束计算机断层扫描的线性头影测量的准确性和可靠性

Accuracy and reliability of linear cephalometric measurements from cone-beam computed tomography scans of a dry human skull.

作者信息

Berco Mauricio, Rigali Paul H, Miner R Matthew, DeLuca Stephelynn, Anderson Nina K, Will Leslie A

机构信息

Department of Developmental Biology, Harvard School of Dental Medicine, Harvard University, Boston, Mass, USA.

出版信息

Am J Orthod Dentofacial Orthop. 2009 Jul;136(1):17.e1-9; discussion 17-8. doi: 10.1016/j.ajodo.2008.08.021.

Abstract

INTRODUCTION

The purpose of this study was to determine the accuracy and reliability of 3-dimensional craniofacial measurements obtained from cone-beam computed tomography (CBCT) scans of a dry human skull.

METHODS

Seventeen landmarks were identified on the skull. CBCT scans were then obtained, with 2 skull orientations during scanning. Twenty-nine interlandmark linear measurements were made directly on the skull and compared with the same measurements made on the CBCT scans. All measurements were made by 2 operators on 4 separate occasions.

RESULTS

The method errors were 0.19, 0.21, and 0.19 mm in the x-, y- and z-axes, respectively. Repeated measures analysis of variance (ANOVA) showed no significant intraoperator or interoperator differences. The mean measurement error was -0.01 mm (SD, 0.129 mm). Five measurement errors were found to be statistically significantly different; however, all measurement errors were below the known voxel size and clinically insignificant. No differences were found in the measurements from the 2 CBCT scan orientations of the skull.

CONCLUSIONS

CBCT allows for clinically accurate and reliable 3-dimensional linear measurements of the craniofacial complex. Moreover, skull orientation during CBCT scanning does not affect the accuracy or the reliability of these measurements.

摘要

引言

本研究的目的是确定从干燥人类头骨的锥形束计算机断层扫描(CBCT)中获得的三维颅面测量的准确性和可靠性。

方法

在头骨上确定了17个标志点。然后进行CBCT扫描,扫描过程中有2种头骨方向。直接在头骨上进行了29个标志点间的线性测量,并与在CBCT扫描上进行的相同测量进行比较。所有测量均由2名操作人员在4个不同场合进行。

结果

在x、y和z轴上的方法误差分别为0.19、0.21和0.19毫米。重复测量方差分析(ANOVA)显示,操作人员内或操作人员间均无显著差异。平均测量误差为-0.01毫米(标准差,0.129毫米)。发现5个测量误差在统计学上有显著差异;然而,所有测量误差均低于已知的体素大小,在临床上无显著意义。在头骨的2种CBCT扫描方向的测量中未发现差异。

结论

CBCT可实现颅面复合体的临床准确且可靠的三维线性测量。此外,CBCT扫描过程中的头骨方向不会影响这些测量的准确性或可靠性。