Berco Mauricio, Rigali Paul H, Miner R Matthew, DeLuca Stephelynn, Anderson Nina K, Will Leslie A
Department of Developmental Biology, Harvard School of Dental Medicine, Harvard University, Boston, Mass, USA.
Am J Orthod Dentofacial Orthop. 2009 Jul;136(1):17.e1-9; discussion 17-8. doi: 10.1016/j.ajodo.2008.08.021.
The purpose of this study was to determine the accuracy and reliability of 3-dimensional craniofacial measurements obtained from cone-beam computed tomography (CBCT) scans of a dry human skull.
Seventeen landmarks were identified on the skull. CBCT scans were then obtained, with 2 skull orientations during scanning. Twenty-nine interlandmark linear measurements were made directly on the skull and compared with the same measurements made on the CBCT scans. All measurements were made by 2 operators on 4 separate occasions.
The method errors were 0.19, 0.21, and 0.19 mm in the x-, y- and z-axes, respectively. Repeated measures analysis of variance (ANOVA) showed no significant intraoperator or interoperator differences. The mean measurement error was -0.01 mm (SD, 0.129 mm). Five measurement errors were found to be statistically significantly different; however, all measurement errors were below the known voxel size and clinically insignificant. No differences were found in the measurements from the 2 CBCT scan orientations of the skull.
CBCT allows for clinically accurate and reliable 3-dimensional linear measurements of the craniofacial complex. Moreover, skull orientation during CBCT scanning does not affect the accuracy or the reliability of these measurements.
本研究的目的是确定从干燥人类头骨的锥形束计算机断层扫描(CBCT)中获得的三维颅面测量的准确性和可靠性。
在头骨上确定了17个标志点。然后进行CBCT扫描,扫描过程中有2种头骨方向。直接在头骨上进行了29个标志点间的线性测量,并与在CBCT扫描上进行的相同测量进行比较。所有测量均由2名操作人员在4个不同场合进行。
在x、y和z轴上的方法误差分别为0.19、0.21和0.19毫米。重复测量方差分析(ANOVA)显示,操作人员内或操作人员间均无显著差异。平均测量误差为-0.01毫米(标准差,0.129毫米)。发现5个测量误差在统计学上有显著差异;然而,所有测量误差均低于已知的体素大小,在临床上无显著意义。在头骨的2种CBCT扫描方向的测量中未发现差异。
CBCT可实现颅面复合体的临床准确且可靠的三维线性测量。此外,CBCT扫描过程中的头骨方向不会影响这些测量的准确性或可靠性。