Westerhoff P, Graichen F, Bender A, Halder A, Beier A, Rohlmann A, Bergmann G
Julius Wolff Institut, Charité-Universitätsmedizin Berlin, D-13353 Berlin, Germany.
J Biomech. 2009 Aug 25;42(12):1840-9. doi: 10.1016/j.jbiomech.2009.05.035. Epub 2009 Jul 30.
Until recently the contact loads acting in the glenohumeral joint have been calculated using musculoskeletal models or measured in vitro. Now, contact forces and moments are measured in vivo using telemeterized shoulder implants. Mean total contact forces from four patients during eight activities of daily living are reported here. Lifting a coffee pot (1.5kg) with straight arm caused an average force of 105.0%BW (%body weight) (range: 90-124.6%BW), while setting down the coffee pot in the same position led to higher forces of 122.9%BW on the average (105.3-153.4%BW). The highest joint contact forces were measured when the straight arm was abducted or elevated by 90 degrees or more, with a weight in the hand. Lifting up 2kg from a board up to head height caused a contact force of 98.3%BW (93-103.6%BW); again, setting it down on the board led to higher forces of 131.5%BW (118.8-144.1%BW). In contrast to previously calculated high loads, the contact force during passive holding of a 10kg weight laterally was only 12.3%BW (9.2-17.9%BW), but when lifting it up to belt height it increased to 91.5%BW (87-95%BW). The moments transferred inside the joint at our patients varied much more than did the forces both inter and intra-individually. Our data suggest that patients with shoulder problems or during the first post-operative weeks after shoulder fractures or joint replacements should avoid certain activities encountered during daily living e.g. lifting or holding a weight with an outstretched arm. Some energy-related optimization criteria used in the literature for analytical musculoskeletal shoulder models must now be reconsidered.
直到最近,肱盂关节的接触负荷一直是通过肌肉骨骼模型计算得出或在体外测量。现在,使用遥测式肩部植入物在体内测量接触力和力矩。本文报告了四名患者在八项日常生活活动中的平均总接触力。伸直手臂拿起一个咖啡壶(1.5千克)时,平均力为105.0%体重(范围:90 - 124.6%体重),而在相同位置放下咖啡壶时,平均力更高,为122.9%体重(105.3 - 153.4%体重)。当伸直手臂外展或抬高90度或更多且手中有重物时,测量到的关节接触力最高。从板子上拿起2千克重物举到头部高度时,接触力为98.3%体重(93 - 103.6%体重);同样,将其放回板子上时,力更高,为131.5%体重(118.8 - 144.1%体重)。与之前计算出的高负荷相反,被动横向握住10千克重物时的接触力仅为12.3%体重(9.2 - 17.9%体重),但将其举到腰带高度时,接触力增加到91.5%体重(87 - 95%体重)。我们患者关节内传递的力矩在个体间和个体内的变化比力大得多。我们的数据表明,肩部有问题的患者或在肩部骨折或关节置换后的术后最初几周内,应避免日常生活中遇到的某些活动,例如伸直手臂举重物或握重物。现在必须重新考虑文献中用于分析性肌肉骨骼肩部模型的一些与能量相关的优化标准。